
41
10.2
AJUNC 3 Board
The AJUNC3 analog board is connected to the digital board, by a flat
cable, on connector JP1. This board contains all the analog and digital
inputs required for the system. The board converts the analog input
signal so the digital board can process it. In addition to this the board
enables the microprocessor (on the digital board) to control all the
digital outputs on the analog board.
The Board’s Connections and Functions:
JP1
Connection between the analog board and the digital board.
JP2
Digital outputs for WATER, EXH, and AIR valves.
Digital inputs for: thermostat, door switch, float switch and
analog input of the electrode.
JP3
Power connection to the board: 5V & 12V.
JP4
Control of the SSR controlling the heating elements.
JP6
MPX2200AP pressure transducer input.
JP10 Voltage output to the fan.
JP11 PT100 temperature sensor input.
JP12 Voltage output to DRY valve (EA, EKA only).
JP13 Communication
port
JP14 Connection to TEST BOARD.
JP15 Connection to drying-pump control (EA. EKA).
Potentiometers
By setting the potentiometers, the input signal is adjusted to the signal
required to be processed by the digital board.
POT1 Setting “Gain Pressure”.
POT4 Setting “Zero Temp”.
POT5 Seting “Gain Temp”.
Содержание 3850 ELVG-D
Страница 43: ...42 AJUNC 3 BOARD...
Страница 53: ...52 VESSEL ASSEMBLY...
Страница 55: ...54 OUTER CABINET ASSEMBLY...
Страница 59: ...58 PIPING DIAGRAM FOR MODELS 1730 E AND EK...
Страница 60: ...59 PIPING DIAGRAM FOR MODELS E AND EK EXCEPT 1730...
Страница 61: ...60 PIPING DIAGRAM FOR MODELS EA AND EKA...
Страница 62: ...61 ELECTRICAL DIAGRAM FOR MODELS 1730 E EK...
Страница 63: ...62 ELECTRICAL DIAGRAM FOR MODELS 2340 2540 E EK...
Страница 64: ...63 ELECTRICAL DIAGRAM FOR MODELS 2340 2540 EA EKA...
Страница 65: ...64 ELECTRICAL DIAGRAM FOR MODEL 3140 E...
Страница 66: ...65 ELECTRICAL DIAGRAM FOR MODELS 3850 3870 E EK...
Страница 67: ...66 ELECTRICAL DIAGRAM FOR MODELS 3850 3870 EA EKA...