27
7.13.2 DPU30
Printer
Caution!
Before starting, disconnect the instrument from the power
source.
1.
Remove the autoclave cover (see para. 7.5
“
Dismantling
the Outer Covers of the Autoclave
”
).
2.
Disconnect the connector of the cable (1) connecting the
electronic box to the printer (2).
3.
Unscrew the two screws (3) attaching the fastening
bracket (4) to the printer.
4.
Remove the printer.
5.
Set the dip switches on the new printer, located on the
back side of the printer, as follows:
6.
Insert the new printer into its frame.
7.
Assemble the fastening bracket to the printer with the
two screws (2) and verify that it is placed firmly.
8.
Connect the (1) cable to the printer.
9.
Verify connection of power by performing a self-test.
10.
Run a cycle and verify that the printer operates correctly.
1
3
2
4
No.
Cat. No.
Description
1 CTP201-0127 Cable, Printer, DPU-30, 30cm, 34p
2 THE002-0022 Printer,
DPU-30, Seiko
3
BOL190-0144 Screw, printer tightening
4
Supplied with
the printer
Fastening bracket
5, 6, 7, 8
down
1, 2, 3, 4,
up
9
up
10
down
Содержание 3850 ELVG-D
Страница 43: ...42 AJUNC 3 BOARD...
Страница 53: ...52 VESSEL ASSEMBLY...
Страница 55: ...54 OUTER CABINET ASSEMBLY...
Страница 59: ...58 PIPING DIAGRAM FOR MODELS 1730 E AND EK...
Страница 60: ...59 PIPING DIAGRAM FOR MODELS E AND EK EXCEPT 1730...
Страница 61: ...60 PIPING DIAGRAM FOR MODELS EA AND EKA...
Страница 62: ...61 ELECTRICAL DIAGRAM FOR MODELS 1730 E EK...
Страница 63: ...62 ELECTRICAL DIAGRAM FOR MODELS 2340 2540 E EK...
Страница 64: ...63 ELECTRICAL DIAGRAM FOR MODELS 2340 2540 EA EKA...
Страница 65: ...64 ELECTRICAL DIAGRAM FOR MODEL 3140 E...
Страница 66: ...65 ELECTRICAL DIAGRAM FOR MODELS 3850 3870 E EK...
Страница 67: ...66 ELECTRICAL DIAGRAM FOR MODELS 3850 3870 EA EKA...