34
7.20
Replacing the Electronics Box
Caution
Ensure the electrical power cord is disconnected!
1. Allow the instrument to cool before removing the outer covers.
2. Remove the autoclave cover (see para. 7.5
“
Dismantling the Outer
Covers of the Autoclave
”
).
3. After the cover is off, remove the two screws that connected the
box to the autoclave base (1).
4. Remove the top screw which is connected the box to the upper
panel frame.
5. Disconnect the connectors No.6, 8 and cable shoe no.7.
6. Remove the PVC tube (5) from the pressure transducer (4). Pull up
the electronic box and replace with a new one.