Installation, Maintenance and Inspection

The  owner  must  ensure  that  all  maintenance,  inspection  and  installation work  is  undertaken  by 

 

authorised and duly qualified personnel who have also studied this manual.

• 

Sensors must always by isolated before starting any work.

• 

Please be aware that the connected sensor(s) and the associated analyser may be controlling 
chemical dosing and as such shutting down the unit without due regard to the systems it is 

 

controlling can lead to chemical release.

Impermissible Modes of Operation

Usage other than as described in this manual will lead to the immediate cancellation of the warranty 
and any other manufacturer’s liability.

Unauthorised Modification

Usage other than as described in this manual will lead to the immediate cancellation of the warranty 
and any other manufacturer’s liability.

Chemicals

Should the sensor(s) be in contact with hazardous chemicals, great care must be taken when handling 
them.

CAUTION

When handling this equipment, the accident prevention regulations applicable on site must be 

 

observed and the specified personal protective equipment worn.

 

PPE: examples of protective clothing, gloves and goggles.

DANGER  

Fire hazard. No parts are suitable for use in a hazardous rated area. 

IMPORTANT

Please unpack the equipment and ordered accessories carefully in order not to miss small parts. 

 

Immediately  compare  the  scope  of  delivery  to  the  delivery  note.  If  there  are  any  discrepancies, 

 

contact your Turtle Tough Representative.

Sensors

Health and Safety

Before making or breaking any electrical or signal connections, ensure that the instrument is isolated 
from the electrical supply. When handling the sensor please wear the appropriate PPE.

Environmental Considerations

Please use this product in a manner sensitive to the environment and at the end of its life dispose 

 

or recycle in a manner appropriate at that time that is in compliance with local regulations. 

Disconnect 

 

electrical supply 

before working 

 

on this equipment

Appropiate 

 

PPE to be 

 

worn when 

 

in contact 

 

with the 

 

sensor

Take care 

 

when

removing 

 

the sensor

Important

Danger

Caution

Warning

Содержание TT-MA-DSS

Страница 1: ...TURTLE TOUGH 1 OPERATION MANUAL TT MA DSS ANALYSER pH ORP Dissolved Oxygen ISE Conductivity EC Smart Digital Sensors TURTLE TOUGH turtletoughsensors com D S S...

Страница 2: ...ors 10 Field Commisioning of Transmitter 11 Wiring Schematic 11 Menu Navigation 12 Function and Programming 13 Parameters Explained 14 Display Features 15 Warranty 17 Return Goods 19 Support 20 Scalin...

Страница 3: ...heading Safety Precautions the special safety precautions outlined in other sections must also be observed Warnings used in this manual This manual contains vital information relating to the safety o...

Страница 4: ...t It is recommended that the outlet is protected by an RCD Residual Current Detector or is in any event in compliance with all local electrical regulations Do not bury electrical supply cable WARNING...

Страница 5: ...CAUTION When handling this equipment the accident prevention regulations applicable on site must be observed and the specified personal protective equipment worn PPE examples of protective clothing gl...

Страница 6: ...of a Turtle Tough TT MA DSS Display Module to ensure you get the maximum life and performance Through our experience in the world s toughest applications we have developed systems that have been optim...

Страница 7: ...ystem includes the following items 110 240V AC Power Supply pH ORP Analyser Module DSS Analyser Module Dissolved Oxygen Analyser Module Conductivity Analyser Module Ion Selective Analyser Module Datal...

Страница 8: ...utes pH compensated Total ISE from Free ISE pH analogue inputs 0 4 20mA analogue Modbus output Smart Digital Modbus RTU Sensor Input TT MA DSS interfaces smart digital DSS MODBUS RTU sensors via 100 d...

Страница 9: ...e 35 to 75 C Weight 75 grams 2 64 ounces Dimensions D 58 x W 36 x H 86 mm 2 3 X 1 4 X 3 4 Ce Mark EN61326A Electrical Power supply 24VDC 10 Power Consumption 60mA max when pH ORP ISE DO 80mA max when...

Страница 10: ...for intelligent maintenance planning smart management of sensor installations and stocking Quick connect digital extension cables facilitate consolidation ofvery many DSS sensors outputs into a one pa...

Страница 11: ...t ti io on n R Ra an ng ge e D De ef fa au ul lt t Software Lock On Off On Address on MODbus Off 1 247 Per Order MODbus baudrate 9 600 19 200 Per Order Toggle for Current Loop Type 4 20mA 0 20mA Per O...

Страница 12: ...01 On no changes can be made Set P01 to Off to allow for changes to scaling and configuration If keys are not used for several minutes then software lock resets back On Lock Function IMPORTANT Please...

Страница 13: ...ralow then choices are uS cm M Standard or M for UPW For Sensor Type 6 uS cm PSU TDS or For Sensor Type 7 uS cm M UPW Per Order 12 Conductivity Sensor Cell Constant K Indicates nominal cell constant f...

Страница 14: ...20mA setpoints Appendix provides percentages corresponding with specific engineered units for various sensor Min scaling between low high setpoints 2 full range Excel worksheet to compute setpoints av...

Страница 15: ...ss otherwise indicated below If P10 is ppm then 0 00 to 150 00 ppm units displayed not flashing as 0 00 to 9 99 10 0 99 9 and 100 150 ppm If P10 is Sat then 0 0 1 500 0 percent saturation units displa...

Страница 16: ...own per scheme above If the Up button is pressed then the mA for the current process value and scaling will be shown If Up button is held for 3 to 5 seconds and P11 is PSU TDS or M then reading in con...

Страница 17: ...quate maintenance faulty installations corrosion or the effects of chemical processes is excluded from this warranty coverage Our agents or representatives may provide you with a life expectancy guide...

Страница 18: ...and plugs must NOT be cut removed or modified in any way Sensors contain sensitive internal electronics and our cables and connectors are designed to protect the integrity of these components Any mod...

Страница 19: ...material plus the name and phone number of a contact person that can be reached for further information Clean Statement returned goods that have been in contact with process fluids must be decontamin...

Страница 20: ...to the information provided in our field implementation notes and guidelines The product is fully supported for installation requirements field implementation and programming protocols in direct rela...

Страница 21: ...Range Engineered pH Units RTU Integer 0 00 2 000 0 5 56 1 000 1000 11 11 0 000 2000 16 67 1 000 3000 22 22 2 000 4000 27 78 3 000 5000 33 33 4 000 6000 38 89 5 000 7000 44 44 6 000 8000 50 00 7 000 9...

Страница 22: ...ed pH Units RTU Integer 0 00 400 0 0 5 00 300 0 1000 10 00 200 0 2000 15 00 100 0 3000 20 00 600 0 4000 25 00 500 0 5000 30 00 400 0 6000 35 00 300 0 7000 40 00 200 0 8000 45 00 100 0 9000 50 00 0 0 1...

Страница 23: ...ORP Units RTU Integer 0 00 2 000 0 0 5 00 1 800 0 1000 10 00 1 600 0 2000 15 00 1 400 0 3000 20 00 1 200 0 4000 25 00 1 000 0 5000 30 00 800 0 6000 35 00 600 0 7000 40 00 400 0 8000 45 00 200 0 9000 5...

Страница 24: ...Engineered DO ppm Limits 0 0 00 15 000 150 00 of Full Range Engineered DO ppm Units RTU Integer 0 00 0 00 0 6 67 10 00 1000 13 33 20 00 2000 20 00 30 00 3000 26 67 40 00 4000 33 33 50 00 5000 40 00 60...

Страница 25: ...teger Limits Engineered DO Sat Limits 0 0 0 15 000 1 500 0 of Full Range Engineered DO Sat Units RTU Integer 0 00 0 0 0 6 67 100 0 1000 13 33 200 0 2000 20 00 300 0 3000 26 67 400 0 4000 33 33 500 0 5...

Страница 26: ...s 0 00 2 000 0 19000000 5 56 1 000 1000 1900000 11 11 0 000 2000 190000 16 67 1 000 3000 19000 22 22 2 000 4000 1900 27 78 3 000 5000 190 33 33 4 000 6000 19 38 89 5 000 7000 1 9 44 44 6 000 8000 0 19...

Страница 27: ...f Full Range 0 01 2 000 0 04 0 00 0 00 1 000 00 50 00 0 02 4 000 0 08 0 00 0 00 2 000 00 50 00 0 05 10 000 0 2 0 00 0 00 5 000 00 50 00 0 10 20 000 0 4 0 00 0 00 10 000 00 50 00 0 20 40 000 0 8 0 00 0...

Страница 28: ...e Engineered PSU MOhm Units RTU Integer 0 00 0 000 0 10 00 5 000 5000 20 00 10 000 10000 30 00 15 000 15000 40 00 20 000 20000 50 00 25 000 25000 60 00 30 000 30000 70 00 35 000 35000 80 00 40 000 400...

Страница 29: ...000 37500 80 00 80 000 40000 85 00 85 000 42500 90 00 90 000 45000 95 00 95 000 47500 100 00 100 000 50000 0 00 0 Default Low Setpoint P06 P07 100 00 100 000 Default High Setpoint P08 P09 CHANGING TDS...

Страница 30: ...e with galvanic DO cells as well as active input measurement module types such as the TT MA pH X TT MA ISE X mate with pH ORP ISE sensors with preamplifiers and TT MA CON all contacting conductivity c...

Страница 31: ...R1 is an internal resistor PLC SCADA Power Supply 24V R1 Input Gnd TT MA 6 7 8 9 10 2 Wire Transmitter In Out Note R1 are internal resistors Input Input 24V Gnd PLC SCADA Power Supply Phoenix or Simil...

Страница 32: ...nsmitter If the TT MA transmitter is reset in the field the analogue 4mA trim offset and 20mA trim gain calibration values shown on this label will be restored from the factory calibration and any mod...

Страница 33: ...oks complicated Will it be hard to use this total ISE measurement module While the documentation on the TT MA TOT module is exhaustive and thorough this is primarily for the purpose to allow further c...

Страница 34: ...d Can the TT MA systems be setup such that the top shell does not have to be removed when replacing sensors Can sensors be changed out without having to pull the sensor cables into the enclosure each...

Страница 35: ...or and the TT MA CON conductivity transmitter to ensure that the cell constants match The TT MA DO transmitters can only be used with the standard thickness membrane sensors such as the ASTDO whereas...

Отзывы: