CON in Micro Siemen units
Measurement
Conductivity
Setup Parameter
Configuration
7
N/A
Sensor Type
6 or 7
P18
Default Node
6
P02
Default Baudrate
19,200
P03
Default Output Type
4-20mA
P04
Default Polarity
non-inverted
P05
Default Low Whole
0
P06
Default Low Decimal
0
P07
Default Hi Whole
100
P08
Default Hi Decimal
0
P09
Days to Recalibrate
90
P17
Units for Output
Con
P11
STANDARD RANGE MODE*
- All values are given in microSiemens/cm
Range Scaling Factor 200
P13
P06/P07
P08/P09
Cell Constant P12
Max
Conductivity
Resolution
0/4mA Low
Setpoint
% of
Full Range
20mA High
Setpoint
% of
Full Range
0.01
200
0.004
0.00
0.00%
200.00
100.00%
0.02
400
0.008
0.00
0.00%
400.00
100.00%
0.05
1,000
0.02
0.00
0.00%
1,000.00
100.00%
0.10
2,000
0.04
0.00
0.00%
2,000.00
100.00%
0.20
4,000
0.08
0.00
0.00%
4,000.00
100.00%
0.50
10,000
0.2
0.00
0.00%
10,000.00
100.00%
1.00
20,000
0.4
0.00
0.00%
20,000.00
100.00%
2.00
40,000
0.8
0.00
0.00%
40,000.00
100.00%
3.00
60,000
1.2
0.00
0.00%
60,000.00
100.00%
5.00
100,000
2
0.00
0.00%
100,000.00
100.00%
10.00
200,000
4
0.00
0.00%
200,000.00
100.00%
20.00
400,000
8
0.00
0.00%
400,000.00
100.00%
HIGH RANGE MODE *
- All values are given in microSiemens/cm
Range Scaling Factor 2,000
P13
P06/P07
P08/P09
Cell Constant
P12
Max
Conductivity
Resolution
0/4mA Low
Setpoint
% of
Full Range
20mA High
Setpoint
% of
Full Range
0.01
2,000
0.04
0.00
0.00%
1,000.00
50.00%
0.02
4,000
0.08
0.00
0.00%
2,000.00
50.00%
0.05
10,000
0.2
0.00
0.00%
5,000.00
50.00%
0.10
20,000
0.4
0.00
0.00%
10,000.00
50.00%
0.20
40,000
0.8
0.00
0.00%
20,000.00
50.00%
0.50
100,000
2
0.00
0.00%
50,000.00
50.00%
1.00
200,000
4
0.00
0.00%
100,000.00
50.00%
2.00
400,000
8
0.00
0.00%
200,000.00
50.00%
3.00
600,000
12
0.00
0.00%
300,000.00
50.00%
5.00
1,000,000
20
0.00
0.00%
500,000.00
50.00%
10.00
2,000,000
40
0.00
0.00%
1,000,000.00
50.00%
20.00
4,000,000
80
0.00
0.00%
2,000,000.00
50.00%
ULTRALOW RANGE MODE*
- All values are given in microSiemens/cm
Range Scaling Factor 2
P13
P06/P07
P08/P09
Cell Constant
P12
Max
Conductivity
Resolution
0/4mA Low
Setpoint
% of
Full Range
20mA High
Setpoint
% of
Full Range
0.01
2
0.00004
0.00
0.00%
2.00
100.00%
0.02
4
0.00008
0.00
0.00%
4.00
100.00%
0.05
10
0.0002
0.00
0.00%
10.00
100.00%
0.10
20
0.0004
0.00
0.00%
20.00
100.00%
0.20
40
0.0008
0.00
0.00%
40.00
100.00%
0.50
100
0.002
0.00
0.00%
100.00
100.00%
1.00
200
0.004
0.00
0.00%
200.00
100.00%
2.00
400
0.008
0.00
0.00%
400.00
100.00%
3.00
600
0.012
0.00
0.00%
600.00
100.00%
5.00
1,000
0.02
0.00
0.00%
1,000.00
100.00%
10.00
2,000
0.04
0.00
0.00%
2,000.00
100.00%
20.00
4,000
0.08
0.00
0.00%
4,000.00
100.00%
TURTLE
TOUGH
®
Note 1
: Difference between low & high analog setpoints should be at least 2% of the full range apart
Note 2:
Minimum recommend Scaling is 4.00% of the full range if the low setpoint is 0.00%.
Note 3
: For high range mode the maximum recommended high 20mA setpoint is 50% of full range