Frequently Asked Questions for TT-MA 

pH, ORP, ISE, Dissolved Oxygen & Conductivity  

transmitters & controllers with application notes 

Can I share the 24VDC power supply used to energize other instrumentation or equipment at the same installation site to also 

energize the TT-MA transmitters?

Absolutely not. 

It is altogether critical that the 24VDC power supply used to power the TT-MA transmitters is COMPLETELY separate from all 

other equipment. 

This also includes all other instrumentation as well other heavier equipment such as pumps, motors and so forth. 

This is because the TT-MA measurement module series is a 3-wire transmitter. The 3-wire aspect can be explained as follows: the 

4-20mA analogue current loop output sent from terminal 7 returns back to terminal 8; the +24VDC is connected to terminal 6 while 

the ground DC common from the power supply is shared with the return of the 4-20mA scalable current loop output on terminal 8. In 

this way all of the ground terminal are shared between the current loop output and the DC common amongst all TT-MA measurement 

modules energized from a single 24VDC power supply source. There exists a 3000V opto-coupler isolation between the inputs and 

outputs of the TT-MA transmitter no matter the particular measurement module. The outputs are not, however, isolated from each 

other as the ground terminal is shared in the manner described above. Because of these reasons whatever 24VDC power supply is 

used to energize the TT-MA transmitters should be altogether dedicated to only power these modules only. This dedicated 24VDC 

power supply can either be customer supplied or using the Turtle Tough supplied TT-MA-PS module.

The 35mm DIN-RAIL mountable TT-MA-PS power supply offer a very simple and low-cost solution for the TT-MA transmitters to 

have their own dedicated 24VDC power supply as required. It is also quite compact being only half the width of a standard TT-

MA module and 35mm DIN-RAIL mountable and so readily fitting into any of the enclosure options. When purchased as part of a 

complete DIN-RAIL mounted or enclosure assembled configuration, the TT-MA-PS module will be pre-wired to all of the TT-MA 

transmitter(s) supplied so that only the AC power input for the TT-MA-PS need to be added to have the unit(s) up and working (plus 

connecting the analogue and MODbus outputs as well as the sensors for input as usual). 

The TT-MA-PS is a well-tested proven power supply solution at very many customer installation sites. The issue of having a dedicated 

24VDC power supply for the TT-MA modules is important for both passive input measurement module types such the TT-MA-pH, 

TT-MA-ISE (mate with pH/ORP/ISE sensors without preamplifiers) and TT-MA-DO (mate with galvanic DO cells) as well as active 

input measurement module types such as the TT-MA-pH-X, TT-MA-ISE-X (mate with pH/ORP/ISE sensors with preamplifiers) and 

TT-MA-CON (all contacting conductivity cells). The isolation on the AC/DC transformer of the TT-MA-PS acts to isolate the 24VDC 

power from the rest of the electrical devices at the installation site ensuring that no such potential ground issues occur to the modules 

themselves nor any devices powered from the TT-MA measurement modules such as preamplifiers and contacting conductivity cells. 

This isolation in the TT-MA-PS also serves to ensure that no issues prevent themselves regarding ground on the analogue 4-20mA 

current loop output or RS485 MODbus outputs emanating from the power supply side of the system. It is still possible to have ground 

loop and electrical isolation issues that emanate from the process side of the system, but this is a much more complex question and 

larger in scope than just the TT-MA instrumentation.

The TT-MA-PS power supply has a 500mA max rating at 24VDC (12 Watts) and easily handles up to 8 each TT-MA modules with 

60mA as the absolute max power consumption per TT-MA module unit. The TT-MA-PS is a universal 100 to 240 VAC 50/60 Hz 

power supply module. The agency approvals for the TT-MA-PS module includes CE, CSA & UL for use in safe areas. If you plan to 

install the TT-MA transmitters into a hazardous area please inquire Turtle Tough for special accommodations necessary to support 

these application and installation types. 

Please see the next page for Power wiring connection diagrams.

How do you wire up the analogue output from the TT-MA transmitter to a data acquisition or control system?

Keep in mind that all TT-MA transmitters are 3-wire devices. This means they have an ACTIVE 4-20mA analogue current loop out

-

put, like a 4-wire type device. The data acquisition or control device to which this TT-MA active 4-20mA output is connected should 

passively measure the current. Most PLC have a hardware or software toggle that allows you to select whether the 4-20mA received 

is from a 4-wire (or 3-wire) active type device or else if it is a 2-wire device which must be energized from the PLC power supply.

NEVER apply voltage across terminals 7 & 8 on any TT-MA transmitter! This could happen if a 3-wire type TT-MA transmitters is wired 

as though it were a 2-wire type devices. The result of such an improper wiring would destroy the output circuit with the damage 

not covered under warranty due to abuse/misuse.

The lead pro24VDC power always goes to terminal 6 and the 4-20mA current loop output is always sent from terminal 7. 

The DC common (ground) is shared as terminal 8. The current loop output is sent from terminal 7 and return to terminal 8 (ground 

/ DC common). The TT-MA transmitters are always energized on terminal 6 with the DC ground of the 24VDC power supply (a.k.a. 

rail) always being the (shared) terminal 8.

Содержание TT-MA-DSS

Страница 1: ...TURTLE TOUGH 1 OPERATION MANUAL TT MA DSS ANALYSER pH ORP Dissolved Oxygen ISE Conductivity EC Smart Digital Sensors TURTLE TOUGH turtletoughsensors com D S S...

Страница 2: ...ors 10 Field Commisioning of Transmitter 11 Wiring Schematic 11 Menu Navigation 12 Function and Programming 13 Parameters Explained 14 Display Features 15 Warranty 17 Return Goods 19 Support 20 Scalin...

Страница 3: ...heading Safety Precautions the special safety precautions outlined in other sections must also be observed Warnings used in this manual This manual contains vital information relating to the safety o...

Страница 4: ...t It is recommended that the outlet is protected by an RCD Residual Current Detector or is in any event in compliance with all local electrical regulations Do not bury electrical supply cable WARNING...

Страница 5: ...CAUTION When handling this equipment the accident prevention regulations applicable on site must be observed and the specified personal protective equipment worn PPE examples of protective clothing gl...

Страница 6: ...of a Turtle Tough TT MA DSS Display Module to ensure you get the maximum life and performance Through our experience in the world s toughest applications we have developed systems that have been optim...

Страница 7: ...ystem includes the following items 110 240V AC Power Supply pH ORP Analyser Module DSS Analyser Module Dissolved Oxygen Analyser Module Conductivity Analyser Module Ion Selective Analyser Module Datal...

Страница 8: ...utes pH compensated Total ISE from Free ISE pH analogue inputs 0 4 20mA analogue Modbus output Smart Digital Modbus RTU Sensor Input TT MA DSS interfaces smart digital DSS MODBUS RTU sensors via 100 d...

Страница 9: ...e 35 to 75 C Weight 75 grams 2 64 ounces Dimensions D 58 x W 36 x H 86 mm 2 3 X 1 4 X 3 4 Ce Mark EN61326A Electrical Power supply 24VDC 10 Power Consumption 60mA max when pH ORP ISE DO 80mA max when...

Страница 10: ...for intelligent maintenance planning smart management of sensor installations and stocking Quick connect digital extension cables facilitate consolidation ofvery many DSS sensors outputs into a one pa...

Страница 11: ...t ti io on n R Ra an ng ge e D De ef fa au ul lt t Software Lock On Off On Address on MODbus Off 1 247 Per Order MODbus baudrate 9 600 19 200 Per Order Toggle for Current Loop Type 4 20mA 0 20mA Per O...

Страница 12: ...01 On no changes can be made Set P01 to Off to allow for changes to scaling and configuration If keys are not used for several minutes then software lock resets back On Lock Function IMPORTANT Please...

Страница 13: ...ralow then choices are uS cm M Standard or M for UPW For Sensor Type 6 uS cm PSU TDS or For Sensor Type 7 uS cm M UPW Per Order 12 Conductivity Sensor Cell Constant K Indicates nominal cell constant f...

Страница 14: ...20mA setpoints Appendix provides percentages corresponding with specific engineered units for various sensor Min scaling between low high setpoints 2 full range Excel worksheet to compute setpoints av...

Страница 15: ...ss otherwise indicated below If P10 is ppm then 0 00 to 150 00 ppm units displayed not flashing as 0 00 to 9 99 10 0 99 9 and 100 150 ppm If P10 is Sat then 0 0 1 500 0 percent saturation units displa...

Страница 16: ...own per scheme above If the Up button is pressed then the mA for the current process value and scaling will be shown If Up button is held for 3 to 5 seconds and P11 is PSU TDS or M then reading in con...

Страница 17: ...quate maintenance faulty installations corrosion or the effects of chemical processes is excluded from this warranty coverage Our agents or representatives may provide you with a life expectancy guide...

Страница 18: ...and plugs must NOT be cut removed or modified in any way Sensors contain sensitive internal electronics and our cables and connectors are designed to protect the integrity of these components Any mod...

Страница 19: ...material plus the name and phone number of a contact person that can be reached for further information Clean Statement returned goods that have been in contact with process fluids must be decontamin...

Страница 20: ...to the information provided in our field implementation notes and guidelines The product is fully supported for installation requirements field implementation and programming protocols in direct rela...

Страница 21: ...Range Engineered pH Units RTU Integer 0 00 2 000 0 5 56 1 000 1000 11 11 0 000 2000 16 67 1 000 3000 22 22 2 000 4000 27 78 3 000 5000 33 33 4 000 6000 38 89 5 000 7000 44 44 6 000 8000 50 00 7 000 9...

Страница 22: ...ed pH Units RTU Integer 0 00 400 0 0 5 00 300 0 1000 10 00 200 0 2000 15 00 100 0 3000 20 00 600 0 4000 25 00 500 0 5000 30 00 400 0 6000 35 00 300 0 7000 40 00 200 0 8000 45 00 100 0 9000 50 00 0 0 1...

Страница 23: ...ORP Units RTU Integer 0 00 2 000 0 0 5 00 1 800 0 1000 10 00 1 600 0 2000 15 00 1 400 0 3000 20 00 1 200 0 4000 25 00 1 000 0 5000 30 00 800 0 6000 35 00 600 0 7000 40 00 400 0 8000 45 00 200 0 9000 5...

Страница 24: ...Engineered DO ppm Limits 0 0 00 15 000 150 00 of Full Range Engineered DO ppm Units RTU Integer 0 00 0 00 0 6 67 10 00 1000 13 33 20 00 2000 20 00 30 00 3000 26 67 40 00 4000 33 33 50 00 5000 40 00 60...

Страница 25: ...teger Limits Engineered DO Sat Limits 0 0 0 15 000 1 500 0 of Full Range Engineered DO Sat Units RTU Integer 0 00 0 0 0 6 67 100 0 1000 13 33 200 0 2000 20 00 300 0 3000 26 67 400 0 4000 33 33 500 0 5...

Страница 26: ...s 0 00 2 000 0 19000000 5 56 1 000 1000 1900000 11 11 0 000 2000 190000 16 67 1 000 3000 19000 22 22 2 000 4000 1900 27 78 3 000 5000 190 33 33 4 000 6000 19 38 89 5 000 7000 1 9 44 44 6 000 8000 0 19...

Страница 27: ...f Full Range 0 01 2 000 0 04 0 00 0 00 1 000 00 50 00 0 02 4 000 0 08 0 00 0 00 2 000 00 50 00 0 05 10 000 0 2 0 00 0 00 5 000 00 50 00 0 10 20 000 0 4 0 00 0 00 10 000 00 50 00 0 20 40 000 0 8 0 00 0...

Страница 28: ...e Engineered PSU MOhm Units RTU Integer 0 00 0 000 0 10 00 5 000 5000 20 00 10 000 10000 30 00 15 000 15000 40 00 20 000 20000 50 00 25 000 25000 60 00 30 000 30000 70 00 35 000 35000 80 00 40 000 400...

Страница 29: ...000 37500 80 00 80 000 40000 85 00 85 000 42500 90 00 90 000 45000 95 00 95 000 47500 100 00 100 000 50000 0 00 0 Default Low Setpoint P06 P07 100 00 100 000 Default High Setpoint P08 P09 CHANGING TDS...

Страница 30: ...e with galvanic DO cells as well as active input measurement module types such as the TT MA pH X TT MA ISE X mate with pH ORP ISE sensors with preamplifiers and TT MA CON all contacting conductivity c...

Страница 31: ...R1 is an internal resistor PLC SCADA Power Supply 24V R1 Input Gnd TT MA 6 7 8 9 10 2 Wire Transmitter In Out Note R1 are internal resistors Input Input 24V Gnd PLC SCADA Power Supply Phoenix or Simil...

Страница 32: ...nsmitter If the TT MA transmitter is reset in the field the analogue 4mA trim offset and 20mA trim gain calibration values shown on this label will be restored from the factory calibration and any mod...

Страница 33: ...oks complicated Will it be hard to use this total ISE measurement module While the documentation on the TT MA TOT module is exhaustive and thorough this is primarily for the purpose to allow further c...

Страница 34: ...d Can the TT MA systems be setup such that the top shell does not have to be removed when replacing sensors Can sensors be changed out without having to pull the sensor cables into the enclosure each...

Страница 35: ...or and the TT MA CON conductivity transmitter to ensure that the cell constants match The TT MA DO transmitters can only be used with the standard thickness membrane sensors such as the ASTDO whereas...

Отзывы: