background image

CyanoFluor U

ser’s Manual 

Rev. A 

14 

4  Measurement Parameters

 

 

 
4.1

 

Phycocyanin (PC) to Chlorophyll (CHL) Ratio Table

 

 

PC to CHL ratios indicate what portion of the total phytoplankton population is comprised of 
PC-containing algae (cyanobacteria).  PC changes proportionally with the abundance of 
cyanobacteria and can be tracked using fluorescence of PC.  When cyanobacteria are not 
present in water the fluorescence from PC is very low yielding ratios less than 0.1; when 
cyanobacteria make up 50% of the total phytoplankton population the PC:CHL ratio will be 
0.5 or greater.  The table below is a quick guide that displays empirically determined 
PC:CHL ratios using lab cultures.  The species of cyanobacteria used to generate these 
data was 

Cyanothece ATCC5201

.  Other algae used were 

Dunaliella salina

Tetraselmis 

sp.

, and 

Rhodomonas salina.  

This table is simply a guide; PC:CHL ratios will vary 

depending on algal species, light history, physiology, etc. It is important to note and monitor 
the rate of change for the PC:CHL ratio over time to determine relative changes in the 
abundance of cyanobacteria group.  As the PC:CHL ratio increases so does the chance that 
a harmful algal bloom will occur.   

 
 

 
 
 
 
 
 
 
 

 

  
 
4.2 Turbidity Effects

 

 

Suspended sediments may decrease accuracy in measurements.  Coastal marine 
environments typically have sediment loads between 5 and 10 NTU equivalents.  High 
sediment loads typically around 25 NTU equivalents will have minimal to no impact on the 
measurements (estimates).  At really high sediment loads (>>25 NTU), consider settling 
samples prior to measuring as you will see effects especially at low algal abundances.     
 
Always check PC and CHL RFU values and compare with blank RFU values to determine 
whether there is enough response from both channels to give a good estimate.  As a 
general rule of thumb, response for at least one channel (PC or CHL) should be more than 
3x the blank RFU values.   

 
4.3 DOM Effects   
 

Dissolved organic materials will affect the chlorophyll channel.  If working in areas where 
DOM is high, it is recommended that you turn on your filtrate blank (section 2.4) and use the 
filtrate correction method described in the Quick Start Guide or Section 2.5/2.6 of this 
manual.  This will correct for any interference or error in estimate due to a high background 
fluorescence which may be cause by dissolved organic materials. 

Cyanobacteria portion 

PC:CHL 

100% cyanobacteria 

> 1.00 

 75% cyanobacteria 

0.75 

 50% cyanobacteria 

0.50 

 25% cyanobacteria 

0.25 

  < 10% cyanobacteria 

<  0.15 

Содержание CyanoFluor

Страница 1: ...P N 998 8701 Revision A February 17 2017 TURNER DESIGNS 1995 N 1st Street San Jose CA 95112 Phone 408 749 0994 FAX 408 749 0998...

Страница 2: ...on Check 9 2 4 Filtrate Blank 9 2 5 Sample Analysis Procedure 10 2 6 Filtering Samples for DOM Correction 11 2 7 Data Display 11 2 8 Downloading Data 11 3 Sample Analysis Guidelines 13 4 Measurement P...

Страница 3: ...an end of life product proceed as follows If you purchased your instrument through a Turner Designs Distributor please contact your local representative They will instruct you where to return the end...

Страница 4: ...production potentially leading to cyanobacterial blooms This monitoring effort can be used to predict the onset of HABs and alert users so the proper protocols can be implemented to prevent or reduce...

Страница 5: ...ips includes CyanoFluor inside a storage pouch batteries are installed 1 Glass 10mm x 10mm Square Cuvette 10 Filter Capsules Data Download cable Laminated Quick Start Guide QSG USB Flash Drive with Us...

Страница 6: ...Then wipe the inside area dry with a clean soft towel or tissue followed by a quick wipe using non abrasive material such as Kim Wipes If extra cleaning is needed use a mild detergent to dampen a tow...

Страница 7: ...selection made If SEND CLEAR DATA is selected press ENTER and a submenu will display Users can choose using keys to either send data to a connected computer or clear logged data a marker will indicate...

Страница 8: ...sing the ON OFF button again will turn the unit off or if left idle on the HOME screen for 3 minutes the unit will automatically turn off to save battery power Note CyanoFluor will NOT automatically t...

Страница 9: ...t clean use a Kim Wipe to wipe the sides of the cuvette before inserting it into the sample compartment wipe away any moisture and or fingerprints 2 Insert the standard into the CyanoFluor s sample co...

Страница 10: ...or fingerprints 5 Press the ON OFF button to turn the instrument on and wait for the 5 second countdown 6 Press the READ button you will be prompted to insert your sample and press read again 7 Making...

Страница 11: ...display Measuring Please Wait NOTE See Sample Analysis Procedure section 2 5 step 9 for result displays 2 7 Viewing Data Data can be viewed on the CyanoFluor s display as well as downloaded to a comp...

Страница 12: ...anager window 7 Click the Browse button and select or create a folder in a desired directory and enter a filename 8 Click Save and then click the Start button on the software s screen 9 Press the ON O...

Страница 13: ...ill help flush out any residual and decrease the chance for cross contamination Do not use a cuvette cap as it may cause the cuvette to not seat properly in the sample compartment When handling the cu...

Страница 14: ...chance that a harmful algal bloom will occur 4 2 Turbidity Effects Suspended sediments may decrease accuracy in measurements Coastal marine environments typically have sediment loads between 5 and 10...

Страница 15: ...easure the sample using the CyanoFluor and record the RFU value for chlorophyll 3 Use extraction methods EPA method 445 0 to determine the actual chlorophyll g L concentration of the water sample meas...

Страница 16: ...ort turnerdesigns com 2 Carry out any adjustments or tests as suggested by the Technical Support Department 3 If proper performance is not obtained you will be issued a Return Authorization number RMA...

Страница 17: ...CyanoFluor User s Manual Rev A 17...

Страница 18: ...le batteries Auto Power Off After 3 minutes of inactivity on the HOME screen only Light Source LED Detector Photodiode LCD Display 2 x 16 characters Resolution 12 bits Case Meets IP 67 Standard dustpr...

Страница 19: ...lear the data log When memory is full the instrument will display Memory Full When READ is pressed you will not be allowed to analyze samples until data have been downloaded and cleared from memory Ba...

Отзывы: