Turner Designs CyanoFluor Скачать руководство пользователя страница 10

CyanoFluor U

ser’s Manual 

Rev. A 

10 

3) 

Press ENTER once <ON> is selected and you’ll return to the HOME screen. 

4)  With the FTR Blank ON, you will be prompted to run a filtrate after every measurement 

to correct for interference from DOM.   
 
To deactivate the FTR Blank simply run through steps 1 

– 3, but select <OFF> and 

press ENTER.  When FTR Blank is OFF, you will not be prompted to run a filtrate blank 
after each measurement 

and the PC and CHL Filtrate Blank will be logged as “NA”.   

 

 

2.5 Sample Analysis Procedure 

 

To make a measurement: 

 

1)  Take up a minimum of 40ccs of sample using a 60cc syringe. 
2)  Rinse the glass cuvette at least three times with the sample in the syringe. 
3)  Fill the rinsed cuvette ¾ full with sample from the syringe. 
4)  Use a Kim Wipe to wipe the sides of the cuvette to wipe away any moisture and/or 

fingerprints. 

5)  Press the ON/OFF button to turn the instrument on and wait for the 5-second 

countdown. 

6)  Press the READ button; you will be prompted to insert your sample and press read 

again. 

7)  Making sure not to spill any sample, insert the cuvette into the sample compartment  
8) 

Press READ; the screen will display Measuring Please Wait… 
 

NOTE: If FTR BLANK is ON, you will be prompted to insert a filtrate blank.  See 
section 2.6 

“Filtering Samples for DOM Correction”, step 1. 

 

9)  Results will be displayed as: 

 
 
 
  

 

10) Press the down arrow key to display Raw Fluorescence Values (RFU) for both PC & 

CHL channels: 

 

 
 
 
 

11) Press the down arrow key again to display FTR Blank values for both PC & CHL 

channels if FTR Blank was ON and a filtrate blank was measured 
 

 
 
 
 
 

 

 

PBLK FILT:        25.00 
CBLK FILT:        12.00 

CHL Channel Filtrate Blank RFU 

 

PC Channel Filtrate Blank RFU 

PC(RFU)            125.00 
CHL(RFU)            82.18 

PC Channel Raw Sample RFU 
 CHL Channel Raw Sample RFU 

 

PC:CHL             1.5211 
20:39                  #0001 

PC to Chl Ratio 

Sample Number (0000 

– 1000

Time (24 hour 
clock; HH:MM) 

Содержание CyanoFluor

Страница 1: ...P N 998 8701 Revision A February 17 2017 TURNER DESIGNS 1995 N 1st Street San Jose CA 95112 Phone 408 749 0994 FAX 408 749 0998...

Страница 2: ...on Check 9 2 4 Filtrate Blank 9 2 5 Sample Analysis Procedure 10 2 6 Filtering Samples for DOM Correction 11 2 7 Data Display 11 2 8 Downloading Data 11 3 Sample Analysis Guidelines 13 4 Measurement P...

Страница 3: ...an end of life product proceed as follows If you purchased your instrument through a Turner Designs Distributor please contact your local representative They will instruct you where to return the end...

Страница 4: ...production potentially leading to cyanobacterial blooms This monitoring effort can be used to predict the onset of HABs and alert users so the proper protocols can be implemented to prevent or reduce...

Страница 5: ...ips includes CyanoFluor inside a storage pouch batteries are installed 1 Glass 10mm x 10mm Square Cuvette 10 Filter Capsules Data Download cable Laminated Quick Start Guide QSG USB Flash Drive with Us...

Страница 6: ...Then wipe the inside area dry with a clean soft towel or tissue followed by a quick wipe using non abrasive material such as Kim Wipes If extra cleaning is needed use a mild detergent to dampen a tow...

Страница 7: ...selection made If SEND CLEAR DATA is selected press ENTER and a submenu will display Users can choose using keys to either send data to a connected computer or clear logged data a marker will indicate...

Страница 8: ...sing the ON OFF button again will turn the unit off or if left idle on the HOME screen for 3 minutes the unit will automatically turn off to save battery power Note CyanoFluor will NOT automatically t...

Страница 9: ...t clean use a Kim Wipe to wipe the sides of the cuvette before inserting it into the sample compartment wipe away any moisture and or fingerprints 2 Insert the standard into the CyanoFluor s sample co...

Страница 10: ...or fingerprints 5 Press the ON OFF button to turn the instrument on and wait for the 5 second countdown 6 Press the READ button you will be prompted to insert your sample and press read again 7 Making...

Страница 11: ...display Measuring Please Wait NOTE See Sample Analysis Procedure section 2 5 step 9 for result displays 2 7 Viewing Data Data can be viewed on the CyanoFluor s display as well as downloaded to a comp...

Страница 12: ...anager window 7 Click the Browse button and select or create a folder in a desired directory and enter a filename 8 Click Save and then click the Start button on the software s screen 9 Press the ON O...

Страница 13: ...ill help flush out any residual and decrease the chance for cross contamination Do not use a cuvette cap as it may cause the cuvette to not seat properly in the sample compartment When handling the cu...

Страница 14: ...chance that a harmful algal bloom will occur 4 2 Turbidity Effects Suspended sediments may decrease accuracy in measurements Coastal marine environments typically have sediment loads between 5 and 10...

Страница 15: ...easure the sample using the CyanoFluor and record the RFU value for chlorophyll 3 Use extraction methods EPA method 445 0 to determine the actual chlorophyll g L concentration of the water sample meas...

Страница 16: ...ort turnerdesigns com 2 Carry out any adjustments or tests as suggested by the Technical Support Department 3 If proper performance is not obtained you will be issued a Return Authorization number RMA...

Страница 17: ...CyanoFluor User s Manual Rev A 17...

Страница 18: ...le batteries Auto Power Off After 3 minutes of inactivity on the HOME screen only Light Source LED Detector Photodiode LCD Display 2 x 16 characters Resolution 12 bits Case Meets IP 67 Standard dustpr...

Страница 19: ...lear the data log When memory is full the instrument will display Memory Full When READ is pressed you will not be allowed to analyze samples until data have been downloaded and cleared from memory Ba...

Отзывы: