background image

CyanoFluor U

ser’s Manual 

Rev. A 

2  Instrument Operation and Calibration Check

 

 

 

2.1 Button Identification and Function 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 

 

 

 

FTR BLANK

 

– is used to correct for interference from Dissolved Organic Materials (DOM) in 

systems that have a high DOM concentration.  See section 2.4 for details. 

 
READ

 

– is used to measure a sample; procedure detailed in section 2.5.   

 
ON/OFF

 

– turns the instrument ON or OFF. 

 
ENTER

 

– allows users to proceed to the next screen/menu. 

 

DATA

 

– when the DATA button is pressed users can choose between either 

sending/clearing data or setting date/time.  Use the 



 keys to choose between the two 

options; a marker will indicate the selection made.   
 
If SEND/CLEAR DATA is selected, press ENTER and a submenu will display.  Users 
can choose, using 



 keys, to either send data to a connected computer or clear 

logged data; a marker will indicate the selection made.    

   

o

  If 

SEND DATA

 is selected and ENTER is pressed, the instrument will send all logged 

data out as an ASCII file - see section 2.8. 

o

  If 

CLEAR DATA

 is selected, the following message will display; 

“ENTER TO 

CONFIRM CLEAR DATA

”; when ENTER is pressed all logged data will be deleted. 

 
If SET DATE/TIME is selected, press ENTER and date and time will display on the 
screen as (MM/DD/YYYY) and (hh:mm).  Use the 



 keys to select either parameter; a 

marker will indicate selection. 
 

o

  If 

MM/DD/YYYY

 is selected, press ENTER and you will be allowed to adjust the 

month, date, and year using the 



 keys.  Press ENTER to save the set date.  

Содержание CyanoFluor

Страница 1: ...P N 998 8701 Revision A February 17 2017 TURNER DESIGNS 1995 N 1st Street San Jose CA 95112 Phone 408 749 0994 FAX 408 749 0998...

Страница 2: ...on Check 9 2 4 Filtrate Blank 9 2 5 Sample Analysis Procedure 10 2 6 Filtering Samples for DOM Correction 11 2 7 Data Display 11 2 8 Downloading Data 11 3 Sample Analysis Guidelines 13 4 Measurement P...

Страница 3: ...an end of life product proceed as follows If you purchased your instrument through a Turner Designs Distributor please contact your local representative They will instruct you where to return the end...

Страница 4: ...production potentially leading to cyanobacterial blooms This monitoring effort can be used to predict the onset of HABs and alert users so the proper protocols can be implemented to prevent or reduce...

Страница 5: ...ips includes CyanoFluor inside a storage pouch batteries are installed 1 Glass 10mm x 10mm Square Cuvette 10 Filter Capsules Data Download cable Laminated Quick Start Guide QSG USB Flash Drive with Us...

Страница 6: ...Then wipe the inside area dry with a clean soft towel or tissue followed by a quick wipe using non abrasive material such as Kim Wipes If extra cleaning is needed use a mild detergent to dampen a tow...

Страница 7: ...selection made If SEND CLEAR DATA is selected press ENTER and a submenu will display Users can choose using keys to either send data to a connected computer or clear logged data a marker will indicate...

Страница 8: ...sing the ON OFF button again will turn the unit off or if left idle on the HOME screen for 3 minutes the unit will automatically turn off to save battery power Note CyanoFluor will NOT automatically t...

Страница 9: ...t clean use a Kim Wipe to wipe the sides of the cuvette before inserting it into the sample compartment wipe away any moisture and or fingerprints 2 Insert the standard into the CyanoFluor s sample co...

Страница 10: ...or fingerprints 5 Press the ON OFF button to turn the instrument on and wait for the 5 second countdown 6 Press the READ button you will be prompted to insert your sample and press read again 7 Making...

Страница 11: ...display Measuring Please Wait NOTE See Sample Analysis Procedure section 2 5 step 9 for result displays 2 7 Viewing Data Data can be viewed on the CyanoFluor s display as well as downloaded to a comp...

Страница 12: ...anager window 7 Click the Browse button and select or create a folder in a desired directory and enter a filename 8 Click Save and then click the Start button on the software s screen 9 Press the ON O...

Страница 13: ...ill help flush out any residual and decrease the chance for cross contamination Do not use a cuvette cap as it may cause the cuvette to not seat properly in the sample compartment When handling the cu...

Страница 14: ...chance that a harmful algal bloom will occur 4 2 Turbidity Effects Suspended sediments may decrease accuracy in measurements Coastal marine environments typically have sediment loads between 5 and 10...

Страница 15: ...easure the sample using the CyanoFluor and record the RFU value for chlorophyll 3 Use extraction methods EPA method 445 0 to determine the actual chlorophyll g L concentration of the water sample meas...

Страница 16: ...ort turnerdesigns com 2 Carry out any adjustments or tests as suggested by the Technical Support Department 3 If proper performance is not obtained you will be issued a Return Authorization number RMA...

Страница 17: ...CyanoFluor User s Manual Rev A 17...

Страница 18: ...le batteries Auto Power Off After 3 minutes of inactivity on the HOME screen only Light Source LED Detector Photodiode LCD Display 2 x 16 characters Resolution 12 bits Case Meets IP 67 Standard dustpr...

Страница 19: ...lear the data log When memory is full the instrument will display Memory Full When READ is pressed you will not be allowed to analyze samples until data have been downloaded and cleared from memory Ba...

Отзывы: