CyanoFluor U
ser’s Manual
Rev. A
8
If ESC is pressed prior to saving the set date, the screen is exited and nothing is
saved. Date is logged per sample.
o
If
hh:mm
is selected, press ENTER and you will be allowed to adjust the hour and
minutes using the
keys. Press ENTER to save the set time. If ESC is
pressed prior to saving the set time, the screen is exited and nothing is saved. Time
is logged per sample.
ESC
– exits to the previous or home screen.
CHK STD
– is used to determine calibration/instrument stability. “Cal. Check Passed” or
“Cal. Check Failed” will be displayed after measuring the CyanoFluor Calibration Check
Standard, using the CHK STD button, indicating whether the CyanoFluor
’s calibration
has changed. The Cal Check Standard P/N 8700-951 is required for this check, see
section 2.3 for details.
2.2 Instrument Power Up
Press the ON/OFF button to power on the CyanoFluor. After a 5 second warm up, the
CyanoFluor is ready for operation. Pressing the ON/OFF button again will turn the unit off or
if left idle on the HOME screen for 3 minutes the unit will automatically turn off to save
battery power.
Note: CyanoFluor will NOT automatically turn off if left idle while running a sample or
if on any screen other than the Home Screen.
HOME SCREEN
The home screen displays the
unit’s current date, time, CyanoFluor firmware version and
remaining battery power as a percentage. These parameters will be displayed after
powering the instrument on and warm up has completed as well as every time the ESC
button is pressed to exit to the home screen.
Battery power is checked whenever users turn the instrument on using the ON/OFF button.
After the 5 second countdown,
-
If the battery power is < 20%, the following warning message will display
“
Battery <20%! Press <ENTER>
” you can continue making measurements
after pressing ENTER, but it is recommended you replace the batteries as
soon as possible.
-
If the battery power is < 10%, the following message will display
“
Battery
Low Pwr! Replace Battery!
” you will not be able to make any
measurements until batteries are replaced.
MM/DD/YYYY HH:MM
FW1.00 Batt: 100%