When TurnAid T4 is connected to the power supply, it is secured against unintended
movement, as the TurnAid T4 will only start by double clicking (2 long pushes)
.
Unauthorised persons such as visitors or the patient cannot start the system. Should the
patient accidentally be lying on a hand control, the person cannot activate TurnAid T4. Lying
on a hand control gives one long push and this is insufficient to start the system! Electrical
functions were secured against unintended operation as defined double clicking on the ON
button is ALWAYS needed to activate the system. Furthermore, the system has a built-in
timer which will turn off the system after 90 seconds.
•
If foot controls are installed on the Turn Aid T4, these will be placed under the bed out of
reach of unauthorised persons.
•
The ON button works as safety stop if here should be a malfunction of the buttons of hand
control, or if someone accidentally presses a foot control switch whilst the system is
connected to the power supply (so the motor rolls will be activated), the activated function
can always be stopped if you push one of these buttons for more than 2 seconds. Then the
motor rolls cannot be operated until you start again as usually with 2 pushes of
.
⚠
TurnAid T4 is an auxiliary for the transfer of a patient. Observance of safety
regulations in connection with the use of the system rests with the buyer.
20.2
Safety
⚠
NEVER use TurnAid T4 otherwise than stated in this instruction manual.
⚠
NEVER leave TurnAid T4 in
"working-position"
when a patient is in the bed.
⚠
ONLY use TurnAid T4 when correctly mounted.
⚠
Keep the hand control out of the patient’s
reach.
⚠
Warning: Risk of foot entrapment when used on a low bed.
⚠
Warning: Materials used on the TurnAid could cause allergies.
⚠
ALWAYS disconnect the system from the power supply if you have to clean, service, repair or
dismount the system. No parts must be services with patient in the system.
⚠
Frequently check that the system
’
s electric wires are intact and ok. All electric parts have been
shielded and double-isolated. Movable parts have been secured with a spring device to eliminate
the danger of clipping/squeezing.
⚠
When setting the 3 TurnAid T4 positions ALWAYS pay attention and care to:
o
Make sure th
e patient’s fingers, arms, legs, hair and any connected equipment are
not caught during the setting of the system.
o
Before and after setting ALWAYS check for clearance around the patient.
o
ALWAYS check that the release handles are locked correctly after setting.