![TurnAid T4 Скачать руководство пользователя страница 28](http://html.mh-extra.com/html/turnaid/t4/t4_instruction-manual_1182044028.webp)
ORIGINAL
page 28
We recommend that you use a transport/mounting trolley when transporting the system. If the
system has to go upstairs, all lifts, where possible, should be made by 2 persons.
8.
Assembly prior to use
Prior to use please check that the bed is suited for TurnAid T4. Installation of TurnAid T4 must
be carried out by a competent person.
8.1
Requirements for the bed
The bed must comply with the standard EN 60601-2-52.
Demands of the bed
manufacturer:
NEVER exceed the
bed manufacturer’s
guidance for max. user load
and max. user weight when TurnAid T4 has been mounted.
Weight TurnAid T4:
48,5 kg.
Calculation of user weight: With a bed of SWL (safe working load) of 350kg, the maximum
patient weight will reduce due to the weight of the TurnAid device.
Please refer to the bed instruction manual for information on the
weight capacity.
8.2
Assembly
Tools necessary for assembly:
•
8 mm Allen HEX wrench
•
4 mm Allen HEX wrench
•
3 mm Allen HEX wrench
•
Folding ruler or measuring
tape
Measure the height from the
bed’s side
-frame bar to the
top of the mattress.
Take off the mattress and
existing bed rails to make
room for TurnAid T4.