4.2
Compatibility between the bed and the mattress
👉
Observe the following requirements for bed and mattress
:
•
BED: The TurnAid T4 system is designed for mounting on beds with square side bars on the
lying surface between the headboard and footboard.
•
Ensure that the gap between the headboard or footboard and the TurnAid T4 meets the
requirements for gaps to ensure there are no entrapment risks. Here you must observe
a distance of 25 to 60 mm. You can adjust TurnAid T4 length to fit between the
headboard and footboard.
•
You can mount TurnAid T4 on beds with various widths. The only requirement is that both
the slide sheet and the turning sheet must fit the width of the bed.
•
MATTRESS: The mattress must fit the bed and comply with the demands of the bed
manufacturer.
We recommend that you firmly inflate an alternating pressure
mattress before operating the
TurnAid T4. A soft mattress will give an uneven support for the patient and thus increase the
friction onto the mattress. If you inflate the mattress firmly, friction is reduced and turning will
become gentler and more comfortable for the patient. If the mattress is soft, you must ensure
that there is no risk of entrapment due to extra compression at the mattress side.
CHANGE OF MATTRESS: When changing the mattress observe that the height for TurnAid
T4
may
have to be altered
–
see chapter 7.
A too low positioning will make it impossible for the nursing staff to change between
"working-
position"
and
"bed rail- position"
as the roll will go against the mattress.