27
Mode
Setting
(Item setting)
Mode
Setting
(Set Number)
Function setting
This is the setting change method of analog output. Operate following the setting code No.
mentioned on page 17.
Code No.75
『
Analog output switching
』
Code No.79
『
Full scale of analog output
』
Select analog output of instantaneous or
totalizer.
Setting range
:
0 (Instantaneous)
1 (Totalizer
)
During analog output of totalizer, the display
equivalent to the max. value can be specified.
Setting range
:
200 to 999999
Change 0 (Instantaneous) to 1 (Totalizer
)
Set display to 5000 for -29 ( 4 to 20mA output)
rated value, when output is 20 mA
MODE
MODE
MODE
▲
SET
SET
SET
SET
▲
MODE
MODE
To code No.setting
Code
№
79
To measurement
operation
Displays of
current setting
Memory
,
▲
Change of setting
ENTER
1sec.
ENTER
1sec.
ENTER
1sec.
1sec.
ENTER
▲
SET
MODE
MODE
SET
SET
▲
MODE
SET
MODE
ENTER
▲
MODE
To code No.setting
To measurement
operation
Code
№
75
Memory
Change of setting
ENTER
1sec.
ENTER
1sec.
1sec.
ENTER
1sec.
Displays of
current setting