Optional output
(
contd.
)
26
Analog output
Please, refer page 9 for "Wiring to this product" of this manual for the connector arrangement.
The measurement input and the analog output are insulated.
Accuracy
Instantaneous
±
0.1
%
of SPAN at 23
℃
±
5
℃
Totalized
±
0.5
%
of SPAN at 23
℃
±
5
℃
Output cycle
Approx. 100ms
Response speed
Approx. 400ms (In case of out instantaneous display)
Condition rated output: 4 to 20mA,
Instantaneous
:
Instantaneous display cycle 100ms
0
→
100
%
output [4
→
20mA During step output]
Selection output data
Selection in either setting mode or instantaneous / totalizer
can be done.
Output of instantaneous display
:
Output depending on instantaneous input
Output of lower 6 digits totalized display
:
Setting of Max. value
(
200 to 999999
)
Output scaling
(
Totalizer
)
Full scale Setting range 200 to 999999