![TSUJI ELECTRONICS CT08-ER2 Скачать руководство пользователя страница 40](http://html1.mh-extra.com/html/tsuji-electronics/ct08-er2/ct08-er2_user-manual_1178497040.webp)
11. Interface with external devices
◆ Regarding to remote controlled input termination switch (2019/10/29 new functionality)
Products that supports remote controlled input termination selection
can use the following commands to control input termination type.
IN?
␣
00
Query termination type for channels 0 to 7.
IN?
␣
08
Query termination type for channels 8 to 15.
THI
␣
ALL
Switch all channels to TTL Hi-Z termination.
THI
␣
x
Switch channel x (where x=0,2,4,…,14) to TTL Hi-Z termination.
(Channel x+1's setting will follow that on channel x)
T50
␣
ALL
Switch all channels to TTL 50 Ohm termination.
T50
␣
x
Switch channel x (where x=0,2,4,…,14) to TTL 50 Ohm termination.
(Channel x+1's setting will follow that on channel x)
NIM
␣
ALL
Switch all channels to NIM termination.
NIM
␣
x
Switch channel x (where x=0,2,4,…,14) to NIM termination.
(Channel x+1's setting will follow that on channel x)
5V
TTL IC
GATE INPUT
GND
LEMO
TTL IC
START, STOP
GND
LEMO
GND
10K
10K