15
Command
$ID8(CHK)(cr)
Syntax
Description
$
First leading code
ID
Address of setting module from 00~FF(HEX)
8
Perform zero calibration
CHK
Checksum
(cr)
Carriage return
Response
!ID(CHK)(cr)
Command valid
?ID(CHK)(cr)
Return not enable calibration before calibration command
Example:
Send command:”$018”,
Response: “!01”
.
*Refer to paragraph 4 Function description before calibration.
5-17. Perform Full Calibration
Command
$ID9(CHK)(cr)
Syntax
Description
$
First leading code
ID
Address of setting module from 00~FF(HEX)
9
Perform full calibration
CHK
Checksum
(cr)
Carriage return
Response
!ID(CHK)(cr)
Command valid
?ID(CHK)(cr)
Return not enable calibration before calibration command
Example:
Send command:”$019”,
Response: “!01”.
*Refer to paragraph 4 Function description before calibration.
5-18. Perform software Reset