Troubleshooting

Indication

Probable cause

Work-around

Solution

All LEDs are OFF 

Even proper voltage is available

Power section of sensor electronic
insert is failed

Sensor electronic insert is needed to be replaced

.

Calibration and settings are all OK 
but mA reading abruptly change or  
chatters continuously

Power supply carrying extra noise and 
capacitance amplifier picking the noise

Ground is not properly connected

Make necessary arrangements to filter the
noise in power-line before being fed to the
device

Provide an exclusive earthing to terminal# 
3, capacitance enclosure earthing screw 
and capacitance probe process connection 
(device mounting screw or flange)

Device contains sufficient filtering of power supply noise 
inside, but sometimes external earth is needed to make 
filters sink the extra power supply noise back to earth, 
connect proper ground.

After calibration current meter showing
21mA and status LED blinking fast 
continuously (Error)

Sense and earth part of probe are shorted
Wrong calibration done at same level

Recalibrate at different level or replace the probe if 
shorted.

Check if probe sense part are shorted with 
ground / PTFE insulation is break and
conductive liquid are entered into it

Maintenance and Spares

Top Cover

Electronic Insert
   connection terminals
   electronic insert fixing screw

   4-way male connector for sensor (probe)
   4-way female connector from sensor (probe)

Cable glands

Thermal spacer (where needed)

Process connection

Stilling well tube

Device enclosure

Shown on the left are various parts of TLC2 capacitance level switch.
Separatable parts are

1. Electronic insert in short called 'electronics'
2. Probe + Enc Cover + Glands collectively called 'mechanical'

For maintenance issues involving replacement of 'electronics', just a
single fixing screw is needed to be released.

Lift the electronics slowly by holding electronics with one hand and
mechanical with other, as wires are connected using rigid 4-way
connectors to it.

Disconnect 4-way connector by holding electronics with one hand and
female of connector by other hand, while the rest of the device is at rest.

Connect the new replaced sensor. 4-way connector is unidirectional and 
only connects in proper direction.

Set the electronics properly to its position.

Match the mounting screw hole of electronics with that of enclosure and 
fix the screw.

For mechanical issues please send the entire device back to Trumen.

tlc2-manual-page04

www.trumen.in

4

Current meter showing 21mA and status 
LED blinking rapidly rapidly by a short
interval (Error)

Current meter shows 20mA very early 
before material level reach at 100% level

High capacitance at probe
Probe is too long for TLC2 in conductive
liquid or water

Wrong Position of Range switch no.5

Check probe length in conductive liquid / 
water if probe length is arround 2500mm or
above

Change the range from 1500 to 4200pF using switch no.5
Throw the switch no.5 to its ON position.

Содержание TLC2

Страница 1: ...ual sensing matters 39 Mangal Nagar Behind Sai Ram Plaza Near Rajiv Gandhi Circle AB Road Indore MP 452 001 India Phone 91 731 497 2065 Trumen Technologies Pvt Ltd Customer Support Phone 91 731 656 24...

Страница 2: ...Operation Matrix TLC SS EIR Page1 Annexure 2 Page1 Page2 Page3 sensing matters Calibration TLC2 EILV ERLV Connection Diagrams TLC2 EILV ERLV Page2 Page3 1 Grounding Load Page2 2 Floating Load Page2 3...

Страница 3: ...ormed with the aid of on site calibration also called wet calibration The probe forms a capacitance with the metalic tank wall The capacitance is sum of three capacitance 4 20 mili Ampere L low Level...

Страница 4: ...y outputs and re transmission over galvanically isolated 4 20mA output SA 80 260VAC 50 60Hz for AC version SD 16 to 32VDC for DC version SPDT 5 A each 24VDC or 220VAC 3 SPDT relays in IP65 max 6 SPDT...

Страница 5: ...uminium housing 5 Make sure process connection is same as that in hopper tank 6 Transmitter probe should never be 6 1 Bent 6 2 Held from thin part 6 3 Cut or machined in any way 6 4 Extended by weldin...

Страница 6: ...way female connector from sensor probe Cable glands Thermal spacer where needed Process connection Stilling well tube Device enclosure Shown on the left are various parts of TLC2 capacitance level sw...

Страница 7: ...d 4 increase 3 Digital Trim decrease Switch 4 Digital Trim increase Switch Switch 4 also act as diagnostic switch 5 Range Switch select for 1500pF to 4200pF 1500pF is required for oil and water up to...

Страница 8: ...SCADA 1 2 3 4 5 6 7 8 9 decrease damping 100 20mA 5V 10V increase U 12 to 60 VDC status to set turn on TLC2 4 20mA o p 1500pF 4200pF Range 1 5V 2 10V Min Supply U 21 5V Vo 0 4mA 1V 2V 3 Voltage Output...

Страница 9: ...witch 3 4 decrease increase ON b Turn 4mA switch ON this sequence is important and watch the mili ammeter till correct trimming of 4mA is done Immediately a Turn switch 3 4 decrease increase OFF b Tur...

Страница 10: ...ill show 20mA Voltmeter will show 5 0V 10 0V LED Indication 4mA or 0 Level 4 mA 20 mA trim decrease trim increase ON 1 2 3 4 When the material in tank is at 0 All switches are OFF LED will blink once...

Страница 11: ...Probe is too long Error is indicated by 21 0mA in loop decrease damping 100 20mA 5V 10V increase 0 4mA 1V 2V 1500pF 4200pF Range 1 2 3 4 5 Status LED Indications TLC2 EILV ERLV Annexure 1 tlc2 manual...

Страница 12: ...ical Connections TLC2 EILV ERLV Annexure 1 tlc2 manual annexure 1 www trumen in Page 6 DC Supply min 12V max 60V Not used TLC2 increase Not used DC Supply min 12V max 60V Voltage output on terminals 2...

Отзывы: