Technical Specification
Features
Specifications
Applications
Speci
fi
cations are subject to change without prior notice
1. Free flowing homogeneous liquids like oil, raw
water, WFI, DM/DI water etc
2. Suitable for top mounting
3. Process temperature max. 200°C
4. Process pressure max. 20 bar
Typical Mountings
1. Fast Switching Response
2. High temperature endurable probes
3. Single sensor allows pump-control & multi-point
switching
4. Easy calibration with or without material
5. Remote electronics with std 10 meters cable length
6. External indication LED available
7. Threaded , Flanged Mountings & TC
8. Electronic Inserts support all requirements
9. Ingress protection : IP 68/65 (as per IS-13947)
10. Ex-proof (Ex d T6 IP-66 IIC)
- Flameproof as per IS/IEC 60079-1:2007
- Weatherproof (IP-66) as per IS/IEC 60529:2001
- Suitable for Gas Group : IIC
- Suitable for Zone 1 & 2 atmospheres
12. Compact size
13. Rigid rod / flexible rope versions
14. No potentiometers - hassle free calibration
compensation against material build-up
Integral Electronics Two-wire Loop Powered
15-60 VDC
4-20mA Loop powered, Error output 21mA/1-5V/2-10V
maximum 475 Ohm @ 24VDC supply
Integral Electronics Three/Four wire (negative common)
15-60 VDC
ModBus-RTU / complementry 0-5V output suitable for >
20K Ohm
Calibration/configuration available through ModBus as well
as without using DIP switches
ICT provides numerical LED indicator, control logic with
relay outputs and re-transmission over galvanically isolated
4-20mA output
SA : 80-260VAC, 50/60Hz for AC version
SD : 16 to 32VDC for DC version
SPDT 5 A each @ 24VDC or 220VAC
(3 SPDT relays in IP65, max 6 SPDT relays in IP40 metal
sheet enclosure)
Contactors with 2NO/2NC rated at
(1, 2 or 3 contactors, only in IP40 metal sheet enclosure)
24V +/- 4V Suitable for maximum 25mA load
4-20mA, Error@21mA, galvanically isolated loop powered
section for use with either integrated ICT Isolated Loop
Supply or any external DC supply within range 16 to
50VDC
Shileded 2 Core PVC cable with 1 to 1.5 mm
2
conductors
cross section
1.8 (non-hygroscopic)
-20°C ... 70°C (-4°F ... 158°F)
-20°C ... 100°C (-4°F ... 212°F)
PTFE Insulation: -30°C ... 250°C (-22°F ... 482°F)
Ceramic Insulation: -30°C ... 600°C (-22°F ... 1,112°F)
(extensions & heat sinks required)
absolute / max. 15 bar (for ceramic insulation : 1 atm)
SS-304, SS-316, SS-316L, PTFE, part ceramic
TC / NPT / BSP 1", 1¼", 1½", 2" etc
Flanged : ANSI/JIS/DIN/ASA/custom
Rigid Rod : 50mm to 3,000mm
Flexible Rope : 100mm to 20,000mm
EIL
Supply
Output
Loop Resistance
EIM
Supply
Interface/Output
ICT specifications
ICT Power Supply
ICT RSx Relay Rating
ICT RKx Relay Rating
ICT Isolated Loop Supply
ICT re-Transmission
ICT to TLC cable
Min. Dielectric Constant
Ambient Temp.
Process Temp.
Extended Process
Temperature
Process Pressure
Wetted Parts
Process Connection
Probe Insertion Length:
ICT remote indicator-controller
using two-core-shielded pvc cable
(1...1.5 mm
2
conductors)
PTFE insulated/bare rope
stilli
ng well tube, flanged m
ounting
inactive
length
(ground)
active l
ength
(sensitive)
tlc2-manual-page02
www.trumen.in
2