56
System Specifications
Result Range: 20-600 mg/dL
Sample Size: 0.5 microliter (0.5 µL)
Sample: Fresh capillary whole blood from the finger and
forearm, or control solution
Test Time: Results in as little as 4 seconds
Result Value: Plasma values
Assay Method: Amperometric
Power Supply: One 3V lithium battery
#CR2032
(non-rechargeable)
Battery Life: Approximately 1000 tests or 1 year
Automatic shut-off: After two minutes of non-use
Weight: 1.66 ounces Size: 3.44” x 2.16” x 0.69”
Memory Size: 1000 glucose results
Operating Range (meter & test strips):
Relative Humidity: 10-90% (Non-condensing)
Temperature: 41-104°F
Hematocrit: 20-70%
Altitude: Up to and including 10,200 feet
Note:
Use within specified environmental conditions only.
Wireless Frequency: 2.4 GHz Band
Contains FCC ID: 2ADDB-NIPRO-01
Device Compatibility: Smartphones running
applications compatible with Bluetooth Glucose Profile
6
Chemical Composition
TRUE METRIX Test Strips: Glucose dehydrogenase-FAD
(Aspergillus species)
, mediators, buffers and stabilizers.
TRUE METRIX Control Solution: Water, d-glucose,
buffers, viscosity enhancing agent, salts, dye and
preservatives.
TMX_AIR_OB_GutsENG_REA4NPD03r5.indd 56
6/4/15 10:48 AM
Содержание METRIX AIR