①
Contact End - Insert into Test Port with contact
blocks facing up.
②
Sample Tip - Bring Tip of test strip to top of sam-
ple drop (blood or control solution) after Drop
Symbol appears in the Display.
Note:
Insert test strip into meter before touching Sample
Tip to blood or control solution drop.
Sample Placement
Correct
Incorrect
• Allow sample drop to be drawn into Sample Tip until test-
ing begins (meter beeps and dashes move across Display)
.
• Do not smear or scrape drop with test strip.
• Do not apply more sample to test strip after testing begins.
• Do not apply blood or control solution to top of test strip.
• Do not insert Sample Tip with sample into Test Port.
May damage meter.
Test Strip
1
2
Top of Test Strip
TMX_AIR_OB_GutsENG_REA4NPD03r5.indd 11
6/4/15 10:47 AM