17
Control Test
We recommend performing Control Tests for practice
before using your meter for the first time to test your
blood.
Control Tests should be performed:
• For practice to ensure your testing technique is good,
• Occasionally as you use a vial of test strips,
• When opening a new vial of test strips,
• If results seem unusually high or low,
• If a vial has been left opened or exposed to humidity or
to extreme heat or cold,
• Whenever a check on performance of the system is
needed,
• If meter damage is suspected (meter was dropped,
crushed, wet, etc.)
Note: It is important to perform Control Tests with more
than one level of control solution. Three levels of
TRUE METRIX Control Solution are available
for Control Tests. Call 1-800-803-6025, available
24 hours a day, 7 days a week, for assistance in
obtaining different levels of control solution.
Caution!
Ranges printed on test strip vial label are for
Control Test results only and are not suggested
levels for your blood glucose. Do not drink
control solution.
TMX_AIR_OB_GutsENG_REA4NPD03r5.indd 17
6/4/15 10:47 AM