To Clean and Disinfect the Meter:
1. Wash hands thoroughly with soap
and water.
2. To Clean: Make sure meter is off
and a test strip is not inserted. With
ONLY PDI Super Sani Cloth Wipes
(EPA* reg. no. 9480-4), rub the
entire outside of the meter using
3 circular wiping motions with
moderate pressure on the front, back,
left side, right side, top and bottom
of the meter. Discard used wipes.
(*Environmental Protection Agency.)
3. To Disinfect: Using fresh wipes,
make sure that all outside surfaces of
the meter remain wet for 2 minutes.
Make sure no liquids enter the Test
Port or an other opening in the
meter.
4. Let meter air dry thoroughly before
using to test.