background image

DL200X operating manual for external sensors

3

EN

���

Battery compartment

���

Battery cover

���

PC connection USB type micro B

���

Network connection RJ45

���

Mount track for fastening

���

LCD display:

���

Measured value row 1

���

Measured value row 2

���

Measured value row 3

������

Display symbol for active acoustic signal

������

Display symbol for active network connection

������

Display symbol for active USB connection

������

Display symbol for power supply over a network

������

Display symbol for power supply over USB

������

Display for battery level

������

Display for mode marker

������

Display for active measured value recording

������

Display for inactive measured value recording

������

Display for date

������

Display for time

������

Mode selection button

������

Data logger for connecting to external sensors

3. Equipment

������

5-pin round connector M12 for connecting up to four digital BUS-compati-
ble sensors:

������

10-pin terminal with two independent input channels for 
connecting up to two analogue sensors:

Pin 2: 
Data B

Pin 1: Te24V (input, ext.
supply for data logger for stationary
operation); also as external power
supply

Pin 3: Terminal A1

Pin 5: Terminal C1

Pin 7: Terminal A0

Pin 9: Terminal C0

Pin 4: Terminal B1

Pin 6: Terminal D1

Pin 8: Terminal B0

Pin 10: Terminal D0

Pin 1: 
Supply sensor 
(output)

Pin 3: Data A

Pin 5: n.c. 

(not connected)

Pin 4: GND (reference
potential)

Pin 2: Terminal GND
(Supply voltage GND)











Содержание DL200X

Страница 1: ...DL200X EN TRT BA DL200X WM 03 EN OPERATING MANUAL DATA LOGGER FOR EXTERNAL SENSORS...

Страница 2: ...ss of the results presented is excluded from any liability or guarantee Liability for damages which have been caused by utilising the presented measured results is strictly excluded 2 Intended use The...

Страница 3: ...ording Display for inactive measured value recording Display for date Display for time Mode selection button Data logger for connecting to external sensors 3 Equipment 5 pin round connector M12 for co...

Страница 4: ...left and the display returns back to the original mode without making changes This function is only available if a BUS sensor has already been registered M61 M81 An 0 M1 M2 M3 M4 M51 M52 M53 M6x M81 A...

Страница 5: ...essary that UDP broadcasts are allowed through the network b Note Based on their concept UDP broadcasts do not operate through a router NAT but instead only within its own network If the data logger i...

Страница 6: ...e BUS sensor to the 5 pin round connector of the data logger see chapter 3 The sensor is supplied with current via the plug By optionally using up to three Y distributors a maxi mum of four digital BU...

Страница 7: ...el 2 has pins 2 4 6 8 and 10 see chapter 3 Both of the input channels support the sensor configurations listed in the following subchapters Example display Registration function M61 selection for regi...

Страница 8: ...r the first input channel Alternatively navigate to the configuration mode M81 AN1 to configure the specification of a sensor for the second input channel The option for configuring the first analogue...

Страница 9: ...1 Further detailed information about configuration of the processing channels in the software is provided in the software manual which you can open by us ing the help function of the SmartGraph softw...

Страница 10: ...on 0 20 mA Current mA cur min max mid Pt100 temperature measurement in 3 cable operation Temperature C cur min max mid Pt100 temperature measurement in 3 cable operation Temperature F cur min max mid...

Страница 11: ...display Additionally a hysteresis can be set by which the measured value must return to the valid range to switch off the alarm Example display for an alarm result of the measuring channel shown in th...

Страница 12: ...itioning for mobile use For mobile measured value recording the data logger can be positioned at any site Observe the permissible ambient conditions for operation see technical data Due to its compact...

Страница 13: ...C Accuracy 0 2 C 0 1 of the measured value Resolution 0 02 C Memory organisation Sensing interval 10 30 s 1 10 12 15 30 min 1 3 6 12 24 h Saving interval 1 10 12 15 30 min 1 3 6 12 24 h Data storage 1...

Страница 14: ...gative polarity approx 0 024 V 0x53 CHANNEL_UNDERRANGE Limit range 0 024 V 0 000 V Displayed 0 000V Normal operation mode 0 000 V 1 000 V Limit range 1 000 V 1 024 V Displayed 1 000V Overvoltage appro...