36 TX6356-UM-EN-01
Press
R
to change from
Latch
and
Auto
.
Press
L
and select
Save
or
Cancel
as
required. Press
R
to confirm the selection
and return to the
Output Setup Menu
.
Checkpoint
A relay configured to
Latch
cannot be reset
until the initiating signal has receded.
Exit
From
Set Relay 1 Mode
or
Set Relay 2
Mode
press
L
, navigate to
Exit
and press
R
to confirm the selection and return to the
Output Setup Menu
.
4.5.2.5 RS485 Data Output Signal
The available menus are as follows:
•
Modbus Address
•
Baud Rate
•
TxOn Delay
•
TxOff Delay
•
Exit
Modbus Address
The
Modbus Address
can be set between
1
and
255
as required.
From the
Output Setup Menu
press
L
,
navigate to
Modbus Address
and press
R
to
enter the
Modbus Address Menu
.
www.trolex.com