7.1 

DC Voltage Measurement:

 

 
 
 
 

 

 
 
 
 
 

 

 

 

  7:

  MEASUREMENT PROCEDURES

 

 

WARNING!

If the magnitude of the voltage to be measured is unknown, always start  

by setting the meter to the highest range, and then to lower ranges,  

until a satisfactory reading is obtained.  

Do not rotate the RANGE switch with the input applied to the meter. 

If the input voltage is higher than 500VDC, or exceeds the limitations  

of a CAT I 600V measurement, do not attempt to measure!

 

 

 

WARNING!

This meter will display voltages that exceed the maximum voltage that 

can safely be applied to the meter. The fact that the meter displays the 

voltages does not mean that it is safe to measure these voltages.

 

7.1.1 

Connect the black test lead to the 

COM

 jack and the red test lead to 

the 

V/

 /mA / F

 jack. 

 
7.1.2 

Set the RANGE switch to a 

V

 

 position. If the magnitude of the 

voltage is unknown, set the RANGE switch to the 500 position.

 

If the 

input voltage is higher than 500VDC, or exceeds the limitations of a 
CAT I 600V measurement, do not attempt to measure! 

  

 

7.1.3 

With power disconnected from the circuit to be measured, connect the 
test probes to the measurement points. Apply power to the circuit. 
The LCD will display the DC voltage. 

 
7.1.4 

If the display indicates overrange, i.e. "1- - -", remove power from the 
circuit, disconnect the test probes, and rotate the RANGE switch to 
the next higher position. Reconnect the test probes, reapply power, 
and read the voltage on the LCD display. If the RANGE switch is 
already at the highest position (i.e. 500 VDC), the input voltage 
exceeds the measurement capability of the meter and should not be 
measured. 

 
7.1.5 

If the displayed value is less than "200" (decimal point not shown), a 
more accurate reading may be obtained by setting the RANGE switch 
to a lower range.  Remove power from the circuit, disconnect the test 

 

13

Содержание 1101-A

Страница 1: ...TRIPLETT MODEL 1101 A 1...

Страница 2: ...Model 1101 A is a 3 1 2 digit handheld Digital Multimeter with a backlit LCD display At its core is an LSI Large Scale Integration integrated circuit which uses dual slope A D conversion for stability...

Страница 3: ...ommendations of any Trade Organizations or Regulatory Agencies whose scope encompasses the use of this meter 2 6 Do not open this meter for maintenance without first disconnecting the test leads from...

Страница 4: ...ltages to the input of the meter which are elevated above the earth ground potential by more than 600VDC AC peak Doing so may damage the meter and or injure the user 2 21 Do not continue to use meter...

Страница 5: ...es These charges may be of lethal potential and may have enough energy to cause damage to the meter or injury to the user These capacitors can store electricity even when the circuitry which they are...

Страница 6: ...the user to misinterpret the hazards present For example the meter may indicate a low voltage when high voltages are actually present 2 46 Avoid usage near strong electrostatic fields high voltage pow...

Страница 7: ...Use caution and follow instructions and warnings to avoid electric shock Failure to do so may result in user injury GROUND The voltage rating associated with this symbol identifies the maximum allowa...

Страница 8: ...git resolution 2000 counts 4 4 Backlit LCD display 4 5 Protective shock absorbing boot 4 6 Diode Test 4 7 Continuity Beeper 4 8 Transistor Gain Test 4 9 Temperature Measurement 4 10 Fused A mA ranges...

Страница 9: ...cations where transient voltages are not present or are of low amplitude Not for use in high energy industrial installation measurement Not for use on circuitry where source capability exceeds 4KW 5 9...

Страница 10: ...0M Overload Protection 500Vrms AC Voltage Range Resolution Accuracy 200V 100mV 500V 1V 1 2 rdg 10 digits Input Impedance All ranges are approx 5M Frequency 40Hz to 400Hz Overload Protection 200mV is 2...

Страница 11: ...s Overload Protection 500Vrms Thermocouple K Type Note The provided thermocouple is only intended to measure Temperatures up to 300 degrees F Exceeding this temperature may damage the probe Diode Test...

Страница 12: ...ight and Hold button 3 Rotary Function Range Selector Switch 4 Transistor Test Jack 5 COM Jack Common black jack for all measurements 6 10A Jack Input for currents from 200mA to 10A 7 V mA F Jack Inpu...

Страница 13: ...magnitude of the voltage is unknown set the RANGE switch to the 500 position If the input voltage is higher than 500VDC or exceeds the limitations of a CAT I 600V measurement do not attempt to measur...

Страница 14: ...meter The fact that the meter displays the voltages does not mean that it is safe to measure these voltages 7 2 1 Connect the black test lead to the COM jack and the red test lead to the V mA F jack 7...

Страница 15: ...rements Do not attempt to measure current in a circuit whose open circuit voltage exceeds 500V 7 3 1 Connect the black test lead to the COM jack Connect the red test lead to the 10A jack unless it is...

Страница 16: ...red test lead to the V mA F jack 7 4 2 Set the RANGE switch to the position that is appropriate for the device or circuit to be measured The LCD display will indicate overrange i e 1 7 4 3 Connect the...

Страница 17: ...he resistance of the device or circuit being measured 7 6 Diode Test WARNING Do not apply voltage or current to the meter when it is set to the Diode Test Continuity Beeper mode 7 6 1 Connect the blac...

Страница 18: ...F jack 7 7 2 Set the RANGE switch to the degree F range 7 7 3 Press the tip of the thermocouple probe against the surface to be measured The tip may also be immersed in non corrosive liquids 7 8 Trans...

Страница 19: ...acklight comes on Release the button The backlight will stay on for about 20 seconds and then turn itself off 8 2 Hold A Hold feature is incorporated into the 1101 A It is a momentary Hold feature tha...

Страница 20: ...ase remove the rubber bumpers that cover the case screws Remove the screws 9 1 2 Gently open case starting at the bottom where the screws were removed 9 1 3 When the back of the case is removed remove...

Страница 21: ...d in the warranty 9 4 Service TRIPLETT PRODUCT RETURN INSTRUCTIONS In the unlikely event that you must return your Triplett equipment for repair the following steps must be taken 1 Call 1 800 TRIPLETT...

Страница 22: ...ole judgment to injure their stability or reliability or which have been subject to misuse abuse misapplication negligence accident or which have had the serial numbers altered defaced or removed Acce...

Страница 23: ...er person is authorized to extend the liability of Triplett Corporation in connection with the sale of its products beyond the terms hereof Triplett Corporation reserves the right to discontinue model...

Страница 24: ...24...

Отзывы: