ZI-9320 

 

 
 

            

Instruction  Manual 

 

         Pocket Digital Multi Meter 

 

 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 

 

 
 
                      

                        

www.zicotech.com

Safety 

International Safety Symbols 

This symbol, adjacent to another symbol or 
terminal, indicates the user must refer to the manual 
for further information. 
 
This symbol, adjacent to a terminal, indicates that, 
under normal use, hazardous voltages may be 
present. 
 
Double insulation 

 

Safety Precautions 

1. 

Improper use of this meter can cause damage, shock, injury or 
death. Read and understand this users manual before 
operating the meter. 

2. 

Make sure any covers or battery doors are properly closed and 
secured. 

3. 

Always disconnect the test leads from any voltage source 
before replacing the battery or fuses. 

4. 

Do not exceed the maximum rated input limits. 

5. 

Use great care when making measurements if the voltages are 
greater than 25VAC rms or 35VDC. These voltages are 
considered a shock hazard. 

6. 

Always discharge capacitors and remove power from the 
device under test before performing Capacitance, Diode, 
Resistance or Continuity tests. 

7. 

Remove the battery from the meter if the meter is to be stored 
for long periods. 

 
 
 
 
 
 
 
Description 

Meter Description 

1. 

3 3/4 Digit (4000 count) 

2. 

RELATIVE button 

3. 

SELECT button 

4. 

DATA HOLD button 

5. 

Hz/DUTY button 

6. 

Function switch 

7. 

Plastic case 

8. 

Test leads 

 

 
 

Specifications 

Electrical Specifications 

Function 

Range 

Accuracy 

DC Voltage 

400.0mV 

(0.7% rdg + 3d) 

   4.000V, 40.00V,  

(1.0% rdg + 3d) 

400.0V, 500V 

(1.3% rdg + 3d) 

AC Voltage 

40-60Hz 

4.000V, 40.00V 

(1.0% rdg + 10d) 

400.0V, 500V 

(2.3% rdg + 5d) 

Resistance 

400.0

4.000k

40.00k

, 400.0k

 

(2.0% rdg + 5d) 

4.000M

 

(5.0% rdg + 5d) 

40.00M

 

(10.0% rdg + 5d) 

Capacitance 

4.000nF 

(5.0% rdg + 30d) 

40.00nF 

400.0nF 

(3.0% rdg + 15d) 

4.000µF, 40.00µF, 

200.0µF 

(10.0% rdg + 15d) 

Frequency 

5.000Hz, 50.00Hz, 

5000.0Hz, 5.000kHz, 
50.00kHz, 500.0kHz, 

10MHz 

(2.0% rdg + 5d) 

Duty Cycle 

0.1-99% 

 

 

 

 

 

 

 

 

 

 

Max input voltage

 

500V AC/DC 

Input Sensitivity,

 

10Vrms min. <9.999KHz 

(Frequency Ranges)

 

40Vrms min. >99.99KHz 

Diode Test

 

Test current 1mA max., open circuit voltage 
of 1.5V typical 

Continuity Check

 

Audible signal if the resistance is < 60

 

Display 

4000 count 3 ¾ digit LCD 

Over range indication

 

LCD displays “OL” 

Polarity

 

Minus (-) sign for negative polarity. 

Low Battery Indication 

“BAT” symbol indicates low battery 
condition. 

Battery

 

 

 

CR2032 3V Lithium 

Operating Temperature

  32

o

F to 104

o

F (0

o

C to 40

o

C) 

Storage Temperature

 

14

o

F to 122

o

F (-10

o

C to 50

o

C) 

Weight

  

 

 

1.7oz (50g) 

Size

  

4.25x2.2x.5” (108x56x11.5mm) 

Standard 

IEC1010  CAT II  500V Pollution degree II, 
CE Approved 

 
 