background image

OPERATION 

— DELAY Setting 

With conventional oscilloscopes, it is not possible to 

observe a trace before triggering signal is developed. 

This oscilloscope has " —  D E L A Y " function to store a 

w a v e f o r m in memory before triggering signal is developed, 

thus permitting a wide variety of applications.  T h e numbers 

indicated by the setting  T R I G G E R POINT  s w i t c h  ( 1 7 ) cor-

respond to the horizontal scale divisions so that trigger 

points can be easily checked on the scope. 

T h e diagram below  s h o w s the  — D E L A Y points indicated by 

the setting  T R I G G E R POINT  s w i t c h  ( 1 7 ) . 

Read-out to Pen Recorder 

This oscilloscope permits any high speed traces to be 

stored in memory and converted into slow speed time base 

which a recorder is able to follow.  T h u s , it is possible to 

record a memory data in a recorder at a speed suitable to 

the recorder's response. 

The standard readout speeds are  1 0 m s / w o r d ,  2 0 m s / w o r d 

and  5 0 m s / w o r d but can be applied to  1 0 0 m s / w o r d , 

2 0 0 m s / w o r d and  5 0 0 m s / w o r d by changing the connec-

tions of the jumper wire in control circuit board (Fig.  2 ) . 

T o record a memory data in a pen recorder, perform the 

following procedures: 

When a trigger point indicated by the setting  T R I G G E R 

POINT  s w i t c h  ( 1 7 ) appears on the scale, the oscilloscope 

provides a negative delay of up to 9 div.  W h e n the 

—  D E L A Y  s w i t c h is  " 6 " position, the memory  w a v e f o r m is 

as  s h o w n in Fig.  1 . 

Note: 

When the  — D E L A Y is set for one-shot signal observation, 

be sure to press the  L E V E L (27) to  N O R M A L  T R I G G E R 

(PUSH) and then press write  S T A R T (16) 

1 . Operate the digital memory and write input signal in 

memory. 

2 . Set the  D I S P L A Y MODE to  D U A L or  M E M O R Y . 

It may be set to  R E A L mode while the  F R E E RUN func-

tion is being used. 

3 . Connect the pen recorder to the  M E M O R Y  O U T FOR 

PEN. 

4 . Set the  C L O C K to  I N T / P E N . 

5 . Depress the  S T A R T , and the memory waveform is read 

out.  T h e readout is completed at  1 0 2 4 words and is set 

in the D/A conversion saturating point.  T o resume the 

readout, depress the  S T A R T once again. 

Memory  w a v e f o r m 

T h e left 6 div  w a v e f o r m is memory  w a v e f o r m  ( 6 0 0 words) 
before triggered.  T h e right 4 div  w a v e f o r m is a memory 
w a v e f o r m  ( 4 0 0  w o r d s ) after triggered. 

T h e  — D E L A Y setting  s w i t c h is used in conjunction  w i t h the 
scale.  W h e n the

 —

  D E L A Y is set to  6 , the trigger point is the  6 t h 

division from the left end of the  s c a l e . 

Trigger point 

1 0 

«  T R I G G E R  P O I N T 6 

S L O P E +-

L E V E L  P U S H 

D I S P L A Y  M O D E  M E M O R Y 

Trigger level 

W a v e f o r m  b e f o r e  t r i g g e r e d 

W a v e f o r m  a f t e r 

t r i g g e r e d 

Fig.

 1 

Содержание MS-1650B

Страница 1: ...MS 1650B DIGITAL MEMORY OSCILLOSCOPE INSTRUCTION MANUAL T R I O...

Страница 2: ...memorized signals on the screen sample intervals have become joined and smooth 9 Real time waveform being displayed on the screen can be stored in the memory thus providing simplified operation 10 Bo...

Страница 3: ...n 1 div of amplitude on t h e C R T L I N E W i t h i n t h e specified p o w e r s u p p l y voltage E X T M o r e t h a n 1 V p p S y n c f r e q u e n c y I N T 2 0 H z 1 0 M H z E X T D C 1 0 M H...

Страница 4: ...e Not more t h a n 5 0 0 ns Input r e s i s t a n c e A p p r o x 3 0 kft 14 pin cable receptacle Data contents 14 pin cable receptacle It Data contents Pin 1 Data L S B Pin 8 M S B Pin 2 B I T 2 Pin...

Страница 5: ...s is used for calibration of the vertical amplifier attenuators and to check the frequency com pensation adjustment of the probes used with the oscilloscope 7 GND TERMINAL Earth terminal of the oscill...

Страница 6: ...ition readout to MEMORY OUT terminal is effective each time S T A R T is depressed For write mode use the SCOPE position The memory data from the initial address to final address are read out once by...

Страница 7: ...eep is started so the triggering level can be adjusted 28 TRIG D Sync indication lamp lights when sync signal is trig gered Check this lamp lights when writing input singal to memory The lamp may ligh...

Страница 8: ...morized signal when connecting MU 1651 memory unit and using as a mutiple trace memory oscilloscope 4 1 D A T A OUT Output terminal for memory data 8 bit BINARY and timing signal 4 2 P O W E R C O N N...

Страница 9: ...e may be checked by feeding input from the CAL 6 terminal Since the calibration voltage is 1 Vp p the waveform becomes 5 div at the 0 2 V position of the VOLTS DIV 8 In measuring DC component set the...

Страница 10: ...t indicated by the setting TRIGGER POINT switch 17 appears on the scale the oscilloscope provides a negative delay of up to 9 div When the DELAY switch is 6 position the memory waveform is as shown in...

Страница 11: ...b this is convenient when it is necessary to observe consecutive phenomena without observer intervention By pulling the DISPLAY TIME knob REAL position the FREE RUN functions Depressing the DISPLAY TI...

Страница 12: ...gnal prior to triggering and should be set to 0 9 div 3 After setting the above switches turn the S T A R T switch to ON The WRITE LED red will light to indicate that the input signal is in standby mo...

Страница 13: ...PLICATIONS OF OSCILLOSCOPE OPERATING Phase Measurement Phase measurements may be made with an oscilloscope Typical applications are in circuits designed to produce a specific phase shift and measureme...

Страница 14: ...ar the MU 1651 memory unit is con nected and its MEMORY OUT is applied to HOR INPUT the phase of two memorized waveforms can be calculated When three M U 1 6 5 K T units are used X Y operation between...

Страница 15: ...signal generator must be extremely flat The oscilloscope vertical input should be set to DC as it will introduce the least distortion especially at low frequencies Because of the harmonic content of...

Страница 16: ...nd that the distorted square wave includes a combination of amplitude and phase distortions In a typical wide band amplifier a square wave check reveals many distortion characteristics of the circuit...

Страница 17: ...Fig 19 indicates low frequency components which have been reduced in amplitude and shifted in phase It will be noted that these examples of low frequency distoriton are characterized by change in shap...

Страница 18: ...oss and phase shift Fig 21 Effect of high frequency boost and good damping F x 1 OUT OF PHASE LAG Fig 18 Tilt resulting from a phase shift of funda mental frequency in a lagging direction Fig 20 Effec...

Страница 19: ...cillation I Low frequency phase shift trace thick ened by hum voltage Fig 2 2 S u m m a r y of waveform analysis for square w a v e testing amplifiers PRECAUTIONS When input signal contains the elemen...

Страница 20: ...ed from the case Caution High voltage of up to 2 0 0 0 V is present at the C R T socket power supply circuit board and FOCUS control To prevent electrical shock be sure to turn off the power when remo...

Страница 21: ...es up or down adjust the VAR BAL VR until it is centered 2 Repeat the above steps so that the trace stays still when the VARIABLE is turned VAR BAL i Fig 2 3 D C B A L a d j u t m e n t Vertical Atten...

Страница 22: ...he four case right side holes with those of retainer plate and fix the re tainer plate with 4 nylon rivets and 4 washers At this time confirm that the retainer plate is installed grommet and insert th...

Страница 23: ...A p r o d u c t o f T R I O K E N W O O D C O R P O R A T I O N 17 5 2 chome Shibuya Shibuya ku Tokyo 150 Japan 3 5 4 0 9 Printed in J a p a n B 5 0 2 9 7 9 0 0 T...

Отзывы: