OPERATION 

OSCILLOSCOPE OPERATION 

Refer to the previous section  " C o n t r o l s on  P a n e l s " . Before 

operating the oscilloscope, set the  s w i t c h e s and controls 

as follows: 

S w i t c h e s and Controls 

Position 

1 .  P O W E R / S C A L E ILLUM 

OFF 

3 .  I N T E N S I T Y 

Center 

(slightly right) 

5.  F O C U S 

Center 

8.  V O L T S / D I V 

2 0 V 

9.  V A R I A B L E 

Full clockwise 

1 0 . 

t POSITION 

Center 

1 2 . 

D I S P L A Y MODE 

R E A L 

1 3 .  A C - G N D - D C 

GND 

2 1 .  S W E E P  T I M E / D I V 

1 ms 

2 2 . 

V A R I A B L E 

Full clockwise 

2 3 . 

<o

 POSITION 

Center 

2 4 . 

FINE  P U L L x 5  M A G 

Center (PUSH) 

2 5 . 

S O U R C E 

INT 

2 6 . 

S L O P E 

2 7 . 

L E V E L 

Center (PULL)  A U T O 

3 1 .  D I S P L A Y  T I M E 

P U S H 

1 . Connect the supplied power cord to the power con-

necter.  T h e oscilloscope is factory adjusted to operate 

on  A C  2 4 0 V . 

2. Turn the  P O W E R  s w i t c h (1) to ON and the  P O W E R lamp 

(2) will light.  T h e n horizontal trace will appear. If 

horizontal trace does not appear at the center of the 

screen, adjust the

 T

 POSITION  ( 1 0 ) . 

3 . Adjust the intensity by the  I N T E N S I T Y  ( 3 ) . If the trace is 

unclear, adjust the  F O C U S  ( 5 ) . 

4 . If the trace is inclined, adjust the  T R A C E  R O T A T I O N  ( 4 ) . 

5. Set the  A C - G N D - D C  ( 1 3 ) to  A C or  D C position and app-

ly input signal to the INPUT  ( 1 4 ) . Turn the  V O L T S / D I V 

(8) clockwise to obtain optimum  w a v e f o r m . 

6. If the  w a v e f o r m is running and not triggered, turn the 

L E V E L  ( 2 7 ) . By pushing the  L E V E L knob, the auto func-

tion is released.  T h e  w a v e f o r m disappears when the 

knob is turned clockwise or counterclockwise and ap-

pears again at the approximate midposition of it. Turn 

the knob until optimum triggering level is obtained. 

7.  W h e n the signal voltage is more than  0 . 0 1 V and 

w a v e f o r m does not appear on the screen, the 

oscilloscope may be checked by feeding input from the 

C A L (6) terminal. Since the calibration voltage is 1 Vp-

p, the  w a v e f o r m becomes 5 div at the  0 . 2 V position of 

the  V O L T S / D I V . 

8. In measuring  D C component, set the  A C - G N D - D C  ( 1 3 ) 

to  D C . If it contains positive ( + } potential, the 

w a v e f o r m moves upward. If negative ( —) potential is 

contained, the  w a v e f o r m moves downward.  T h e zero 

potential can be checked at the GND position. 

DIGITAL MEMORY OPERATION 

Set the  s w i t c h e s and controls  a s follows before using the 

digital memory: 

Other knobs and controls are the same in function  a s those 

of oscilloscope operation. 

S w i t c h e s and Controls 

Position 

1 1 .  M E M O R Y POSITION 

Pull mechanical 

center 

1 2 . 

D I S P L A Y MODE 

R E A L or  D U A L 

1 5 .  C L O C K 

I N T / S C O P E 

1 7 .  T R I G G E R POINT 

2 1 .  S W E E P  T I M E / D I V 

0 . 1 ms — 1 s 

2 7 . 

L E V E L 

N O R M A L (PUSH) 

3 1 .  D I S P L A Y  T I M E 

P U S H 

3 2 . 

PEN  S P E E D 

10  m s ,  2 0  m s , 

5 0 ms/word 

[A] Manual

 Operation 

1 . Operate the oscilloscope and display a triggered signal 

on the screen. 

2. Depress the  W R I T E  S T A R T  ( 1 6 ) and the  W R I T E  ( 1 8 ) 

lamp will light. 

3 .  W h e n the  W R I T E lamp goes off, the  R E A D (19) lamp 

begins to flicker. 

4 . Set the  D I S P L A Y MODE  ( 1 2 ) to  D U A L or  M E M O R Y , and 

memory  w a v e f o r m will be displayed. In the  D U A L mode, 

the real time  w a v e f o r m overlaps  w i t h the memory 

w a v e f o r m .  B y adjusting the

 T

  P O S I T I O N  ( 1 0 ) or 

M E M O R Y POSITION  ( 1 1 ) , both  w a v e f o r m s can be 

observed. 

5.  W h e n observing memory  w a v e f o r m , it is not necessary 

to adjust the  V O L T S / D I V (8) and  S W E E P  T I M E / D I V 

( 2 1 ) .  T h e  V O L T S / D I V does not change the amplitude of 

the memory  w a v e f o r m .  T h e  S W E E P  T I M E / D I V changes 

only the readout speed; the memory  w a v e f o r m remains 

unchanged on the screen. 

[B] Automatic write-in, read-out Operation 

Pull out the  D I S P L A Y  T I M E  ( 3 1 ) forward you and set this 

control at about 9 o'clock position in  F R E E RUN state. 

Refer to  F R E E RUN function on page  1 1 . 

Содержание MS-1650B

Страница 1: ...MS 1650B DIGITAL MEMORY OSCILLOSCOPE INSTRUCTION MANUAL T R I O...

Страница 2: ...memorized signals on the screen sample intervals have become joined and smooth 9 Real time waveform being displayed on the screen can be stored in the memory thus providing simplified operation 10 Bo...

Страница 3: ...n 1 div of amplitude on t h e C R T L I N E W i t h i n t h e specified p o w e r s u p p l y voltage E X T M o r e t h a n 1 V p p S y n c f r e q u e n c y I N T 2 0 H z 1 0 M H z E X T D C 1 0 M H...

Страница 4: ...e Not more t h a n 5 0 0 ns Input r e s i s t a n c e A p p r o x 3 0 kft 14 pin cable receptacle Data contents 14 pin cable receptacle It Data contents Pin 1 Data L S B Pin 8 M S B Pin 2 B I T 2 Pin...

Страница 5: ...s is used for calibration of the vertical amplifier attenuators and to check the frequency com pensation adjustment of the probes used with the oscilloscope 7 GND TERMINAL Earth terminal of the oscill...

Страница 6: ...ition readout to MEMORY OUT terminal is effective each time S T A R T is depressed For write mode use the SCOPE position The memory data from the initial address to final address are read out once by...

Страница 7: ...eep is started so the triggering level can be adjusted 28 TRIG D Sync indication lamp lights when sync signal is trig gered Check this lamp lights when writing input singal to memory The lamp may ligh...

Страница 8: ...morized signal when connecting MU 1651 memory unit and using as a mutiple trace memory oscilloscope 4 1 D A T A OUT Output terminal for memory data 8 bit BINARY and timing signal 4 2 P O W E R C O N N...

Страница 9: ...e may be checked by feeding input from the CAL 6 terminal Since the calibration voltage is 1 Vp p the waveform becomes 5 div at the 0 2 V position of the VOLTS DIV 8 In measuring DC component set the...

Страница 10: ...t indicated by the setting TRIGGER POINT switch 17 appears on the scale the oscilloscope provides a negative delay of up to 9 div When the DELAY switch is 6 position the memory waveform is as shown in...

Страница 11: ...b this is convenient when it is necessary to observe consecutive phenomena without observer intervention By pulling the DISPLAY TIME knob REAL position the FREE RUN functions Depressing the DISPLAY TI...

Страница 12: ...gnal prior to triggering and should be set to 0 9 div 3 After setting the above switches turn the S T A R T switch to ON The WRITE LED red will light to indicate that the input signal is in standby mo...

Страница 13: ...PLICATIONS OF OSCILLOSCOPE OPERATING Phase Measurement Phase measurements may be made with an oscilloscope Typical applications are in circuits designed to produce a specific phase shift and measureme...

Страница 14: ...ar the MU 1651 memory unit is con nected and its MEMORY OUT is applied to HOR INPUT the phase of two memorized waveforms can be calculated When three M U 1 6 5 K T units are used X Y operation between...

Страница 15: ...signal generator must be extremely flat The oscilloscope vertical input should be set to DC as it will introduce the least distortion especially at low frequencies Because of the harmonic content of...

Страница 16: ...nd that the distorted square wave includes a combination of amplitude and phase distortions In a typical wide band amplifier a square wave check reveals many distortion characteristics of the circuit...

Страница 17: ...Fig 19 indicates low frequency components which have been reduced in amplitude and shifted in phase It will be noted that these examples of low frequency distoriton are characterized by change in shap...

Страница 18: ...oss and phase shift Fig 21 Effect of high frequency boost and good damping F x 1 OUT OF PHASE LAG Fig 18 Tilt resulting from a phase shift of funda mental frequency in a lagging direction Fig 20 Effec...

Страница 19: ...cillation I Low frequency phase shift trace thick ened by hum voltage Fig 2 2 S u m m a r y of waveform analysis for square w a v e testing amplifiers PRECAUTIONS When input signal contains the elemen...

Страница 20: ...ed from the case Caution High voltage of up to 2 0 0 0 V is present at the C R T socket power supply circuit board and FOCUS control To prevent electrical shock be sure to turn off the power when remo...

Страница 21: ...es up or down adjust the VAR BAL VR until it is centered 2 Repeat the above steps so that the trace stays still when the VARIABLE is turned VAR BAL i Fig 2 3 D C B A L a d j u t m e n t Vertical Atten...

Страница 22: ...he four case right side holes with those of retainer plate and fix the re tainer plate with 4 nylon rivets and 4 washers At this time confirm that the retainer plate is installed grommet and insert th...

Страница 23: ...A p r o d u c t o f T R I O K E N W O O D C O R P O R A T I O N 17 5 2 chome Shibuya Shibuya ku Tokyo 150 Japan 3 5 4 0 9 Printed in J a p a n B 5 0 2 9 7 9 0 0 T...

Отзывы: