Calibration & Connectivity | TS-700 Series
7
Calibration Switch, TS-700 MS:
Remove the
screws and the back plate of the enclosure. A
white-capped calibration switch is located at the
bottom left corner of the circuit board. Press the
switch fully in and release. This will place the
switch in the up position, allowing access to the
calibration parameters. Loosely reattach the
rear plate.
Calibration Switch, TS-700 SS:
Remove the
screws and the back plate of the enclosure. A
calibration switch is located at the bottom left
corner of the circuit board. Move the switch to
the “On” position, allowing access to the
calibration parameters. Loosely reattach the
rear plate.
Calibration Switch, TS-700 WB:
Disconnect all
cables and cords. To remove the back plate of
the enclosure, place the indicator upside down
(not face down) on a work surface. Locate the
two slotted holes near the connection sockets.
Insert a flat blade screwdriver into each slot
(one at a time), pushing the blade inward while
angled towards the back of the indicator. This
will release the retaining prong and enable
removal of the back plate. Carefully lift the rear
plate.
A white-capped calibration switch is located at
the bottom left corner of the circuit board. Press
the switch fully in and release. This will place
the switch in the up position, allowing access to
the calibration parameters. Loosely reattach the
rear plate.
ALL MODELS:
After enabling access to