Calibration & Connectivity | TS-700 Series
6
Calibrating a TS-700 Series Indicator
NOTICE
: The TS-700 SS/MS/WB are
NTEP Class III Legal for Trade approved
and certified. Unlocking the calibration
parameters is not allowed by persons
other than qualified scale technicians.
ALL MODELS: Place the Indicator in
Calibration Mode and Determine if the
Calibration Parameters are Accessible
1) Power on the indicator.
2) Press the PRINT and HOLD keys. “C” will
display. Release the keys.
3) If the calibration parameters are accessible,
“C01” will display, with the 1 flashing. Proceed
to page 8, “CALIBRATION PROCEDURE”.
NOTE
: If C08 Displays instead of C01, the
calibration switch must be adjusted. Disconnect
the AC adapter and power off the indicator.
DEFAULT SETTINGS CAUTION
Placing the indicator in calibration mode enables
access to the Default Settings Parameter, CO7.
NEVER activate Parameter C07 unless you are a
qualified scale technician,
and:
A) You understand that all custom settings will be
deleted permanently, and
B) You understand how to restore all custom
parameters to the required settings based on the
scale’s specifications and usage.