Calibration & Connectivity | TS-700 Series
15
Error Codes
ERROR CAUSE SOLUTION
uuuuuu
1. Overloaded condition.
2. Wrong connection at
load cell.
3. Load cell has quality
problem.
1. Reduce the weight.
2. Check load cell
connection
3. Inspect load cell.
Check the input and
output.
nnnnnn
1. Calibration is incorrect.
2. Bad connections.
3. Load cell has quality
problem
.
1. Check scale is resisted
or not, foot is kept
level or not.
2. Check load cell
connection.
3. Check input and
output resistance of
load cell
.
ERR1
Overweight calibration
error.
Incorrect weight input or
incorrect weight on
weighing platform.
1. Input weight correctly
during calibration.
2. Place correct weight
on weighing platform.
ERR2
Underweight calibration
error.
Incorrect weight input or
incorrect weight on
weighing platform.
The calibration weights
Minimum is 10% of
Max. cap.
Recommend 60%-
80% of Max. Cap.
ERR3
During calibration, the
input signal is negative.
Check all connections.
Check load cell.
If connections and
load cell are okay,
PCB needs replacing.
ERR4
During calibration, the
signal is unstable.
After the platform is
stable, start
calibration.
ERR5
Circuitry error
Replace the PCB.