Calibration & Connectivity | TS-700 Series         

 

 

 

11 

C06 

Calibration 

0 = skip calibration 

  1 = proceed with calibration  
 
Description:  

With C06 flashing, press the 

 key, then use the up 

arrow key to change the 
parameter setting to 1. 
Press the 

 key to 

proceed. “Span” will briefly 
display, then all digits will 
display. 
Load the weighing platform, 
then adjust the displayed 
digits to match the weight on 
the platform, i.e, 4,000.  
Press the 

 key to 

proceed.  The indicator will 
calibrate, then display 
“CalEnd”. 

 

PRESS THE 

 KEY TO 

ACCEPT THE 
CALIBRATION AND EXIT 
SETTINGS MODE. 

CAUTION -- IMPORTANT NOTICE!! 

After calibrating, be sure to exit settings mode 
without activating parameter C07! 

 
Parameter C07 is the parameter that can DELETE all 
custom settings and reset the indicator to factory 
settings. 

 

C07 
CAUTION!!

 

FACTORY 
RESET… 
Erases all 
custom 
settings! 

0 = skip reset to factory 

defaults 

1 = proceed with reset to 

defaults  

 

Description: 
To reset to factory defaults, 
change the 0 setting to 1 and 
press the 

 key.

 

Содержание TS 700-WB

Страница 1: ...ocument supplements the User s Guide www trinerscale com Version 4 0 Calibration Connectivity TS 700 Series Digital Indicators Full Function Advanced Function Digital Indicator TS 700 MS TS 700 SS TS...

Страница 2: ...Triner Scale Mfg Co Inc 8411 Hacks Cross Road Olive Branch MS 38654 800 238 0152 www trinerscale com 2013 Triner Scale All rights reserved...

Страница 3: ...S 700 Series Indicator 6 Data Output Formats 12 Error Codes 15 Factory Default Settings 16 CALIBRATION WARNING Calibration AND inspection of calibration properties is prohibited unless done so by a qu...

Страница 4: ...tion temp 10 C 40 C Operation humidity 90 RH Storage temp 40 C 70 C NTEP CLASS III COC 09 080 KEYPAD KEYS FUNCTIONS requires adjustment of parameter settings Hold Hold weight on scale or Hold peak wei...

Страница 5: ...tions The indicator can connect with a maximum of eight 8 350 load cells 4 wire or 6 wire configurations Excitation voltage for the load cell is 5VDC the maximum output current is 120mA TS 700MS WB Lo...

Страница 6: ...S 700 MS The RS232 DB9 serial port can be utilized to connect the indicator to a peripheral device such as a printer or remote display or to interface the indicator with a PC NOTE PC interface require...

Страница 7: ...rface the indicator with a PC NOTE PC interface requires Triner DataLog PC software to be installed on the PC TS 700 SS RS 232 Terminal Block TS 700 WB The RS232 3 pin quick connect serial port can be...

Страница 8: ...he keys 3 If the calibration parameters are accessible C01 will display with the 1 flashing Proceed to page 8 CALIBRATION PROCEDURE NOTE If C08 Displays instead of C01 the calibration switch must be a...

Страница 9: ...bration parameters Loosely reattach the rear plate Calibration Switch TS 700 WB Disconnect all cables and cords To remove the back plate of the enclosure place the indicator upside down not face down...

Страница 10: ...rease the value of the flashing digit or to navigate up and down a list of preset options The and arrow keys are used to move the active digit over to the left or to the right The key and key can be u...

Страница 11: ...the parameter setting as needed Press the key to accept the setting and continue on to the next parameter C02 Decimal location 0 none 1 one decimal place 2 two decimal places 3 three decimal places De...

Страница 12: ...ion With C04 flashing press the key then use the left and right arrow keys to select the active digit then use the up and down keys to adjust the digits to the desired capacity i e 5 000 Press the key...

Страница 13: ...platform i e 4 000 Press the key to proceed The indicator will calibrate then display CalEnd PRESS THE KEY TO ACCEPT THE CALIBRATION AND EXIT SETTINGS MODE CAUTION IMPORTANT NOTICE After calibrating...

Страница 14: ...2 Decimal point position 1 0 0 XXXXXX0 0 1 0 XXXXXXX 1 1 0 XXXXX X 0 0 1 XXXX XX 1 0 1 XXX XXX Bits 3 4 Division 0 1 X1 1 0 X2 State B BITS FUNCTION Bits0 Gross 0 net 1 Bits1 Symbol positive 0 negati...

Страница 15: ...splay 1 Bit 5 Constant 1 Bit 6 Constant 0 PC COMPUTER CONTINUOUS SENDING FORMAT CR L F S 1 S 2 S 3 Data S 4 S1 Weight status ST standstill US not standstill OL overload S2 Weight mode GS gross mode NT...

Страница 16: ...ollows Command Name Function T TARE Save and clear tare Z ZERO Zero gross weight P PRINT Print the weight R G W N W Read gross weight or net weight C Kg lb Kg lb conversion G G W Check gross weight at...

Страница 17: ...esistance of load cell ERR1 Overweight calibration error Incorrect weight input or incorrect weight on weighing platform 1 Input weight correctly during calibration 2 Place correct weight on weighing...

Страница 18: ...ty calibration 0 C07 Restore default 0 C08 Warning tone 1 C09 Power off automatically 0 C10 Power saving mode 0 C11 Hold function 0 C12 Disable kg lb conversion 1 C13 Upper limit alarm 000000 C14 Unde...

Страница 19: ...tandstill time 1 C27 Standstill range 2 C28 Dynamic filter 0 C29 Noise filter 2 C30 Print format 0 C31 Analog signal options 1 C32 4 20mA testing 4 C33 Relay output setting 1 C34 Reserved 0 C35 Reserv...

Страница 20: ...Triner Scale Manufacturing Co Inc 8411 Hacks Cross Road Olive Branch MS 38654 800 238 0152 www trinerscale com...

Отзывы: