Version: TCM_0450_E80_M_EN_150209_E007
64
Manual Operation
HART
®
For the HART
®
interface special local settings are not available. All configurations are done directly via the
interface.
The HART
®
interface is just activated, and the display returns to the SETUP IN/OUTPUTS menu.
Foundation Fieldbus
®
For the Foundation Fieldbus
®
interface special local settings are not available. All configurations are done
directly via the interface.
The Foundation Fieldbus
®
interface is just activated and the display returns to the SETUP IN/OUTPUTS menu.
4.5.9.
SETUP DATA CONFIGURATION Menu
In the submenu SETUP/DATA CONFIG the current settings can be stored to the backup memory and the stored
settings can be recalled
SETUP MENU
DATA CONFIG
UP DOWN EXIT
P
DATA CONFIG MENU
SAVE DATA
UP DOWN EXIT
Use the keys
“UP” and “DOWN” to select the desired submenu and confirm with “P” or skip with “EXIT”.
4.5.9.1.
SAVE DATA Menu
In the submenu SAVE DATA the current settings can be stored in the backup memory.
At each power on the TCE 8000 compares the content of the setup memory and the backup memory. If the data
in those two memories are different, the TCE 8000 gives out a warning. To avoid this warning, it is
recommended to make a backup as soon as the new settings are proven to be okay.
DATA CONFIG MENU
SAVE DATA
UP DOWN EXIT
P
READY TO SAVE DATA
START EXIT
Start the backup process with
“START” or skip with “EXIT”.
If START is pressed, the display shows for some seconds
MEMORY
ACCESS
After that for about two seconds
READY
The display returns to the DATA CONFIG menu.