4.5.
Control Mode
In the control mode the TCE 8000 can be adapted to the individual application. As unintentional changes of the
settings might cause problems, some submenus are password protected. Additionally, using the
“SERVICE”
menu, a global access code can be set that locks the TRICOR Coriolis Mass Flow Meter completely.
To enter the control mode proceed as follows:
Press
“P” for about three seconds
If a global access code is set the display shows
ENTER P-ACCESS CODE
0000
LEFT UP EXIT
Change the indicated number with
“LEFT” and “UP” to the defined code and confirm with “P”.
If a wrong code is entered, the display shows
“ERROR” for about two seconds and then returns to the
measurement mode.
When the correct code is entered the display shows
MAIN MENU
ZERO OFFSET
UP DOWN EXIT
With the keys
“UP” and “DOWN” you can scroll through the main list.
Select the desired submenu and confirm with
“P”.
NOTE:
With the
“Custody Transfer” option, no settings can be configured when the housing is closed since the P button
is deactivated.
4.5.1.
Function of the Pushbuttons
In the setup menu some pushbuttons have changing functions, indicated in the display above the pushbutton:
P
Confirms the selection in a list or any kind of inputs
Reset
Performs the indicated function
Display
Performs the indicated function.
Info (Exit)
Performs the indicated function.
In most cases exits the current menu point without altering the original value