Version: TCM_0450_E80_M_EN_150209_E007
39
Manual Operation
Navigating between log entries
Press the
“Display” button to show the previous event (log entry) in chronological order:
** LOG DISPLAY **
LOG#: 1/2 INFO
TIMESTAMP: 0,005s
POWER ON SEQUENCE
Press the
“Info” button to navigate to the previous log entry. You can repeat this until you get to the last log
entry.
Info menu
When the last log entry is displayed (for instance LOG#: 5/5 INFO), pressing the
“Info” button again opens the
Info menu that lists the following 7 internal device parameters:
Code
Value
SA
Sensor voltage A in mV
SB
Sensor voltage B in mV
DR
Drive current in mA
PT
Resistance value of the temperature sensor in
Ω
FRE
Oscillating frequency in Hz
ZP
Zero point offset in µs
TSF
Filtered time shift in µs
Tab. 4: List of service parameters
Press
“Info” again to get general information about the TCM:
Code
Value
TYPE:
Sensor type (TCM*)
SER.:
Serial number
SW1:
SW version main board
SW2:
SW version display
COMM:
Active interface:
Modbus/baud rate/address
HART
®
/address
FF/address
CODE:
SW option code
Tab. 5: List of TCM device info
By pressing
“Info” you can toggle between those info views. To return to the normal operation screen, press
“DISPLAY”.