Table 11.1
Quick Troubleshooting Procedures
Symptom
Probable Cause
Remedy
On/off power switch is in on
position. Unit does not operate.
AC line cord not plugged in or
damaged.
AC line fuse blown.
Test cord for correct seating in
socket or for shorts.
Remove fuse and determine if
blown.
Power is on. Touch panel down
not functioning.
Various malfunctions.
Have unit serviced.
No fixer dispensing.
Pumps not primed.
Pump not working.
Hold fixer button until fixer starts
to flow or go to Diagnostics tab.
Use Primer bulb
Have unit serviced.
No suction at drain head.
Broken suction line or vacuum
pump.
Suction o-ring missing.
Have unit serviced.
Remove magnet cover and
replace small o-ring
Sample not dispensing.
Pump Tubing Pinched
Peristaltic pump tube leaking.
O-ring not sealing sample tubing.
Sample head tube seal leaking.
Peristaltic pump not turning.
See Maintenance Section 11.8
Have unit serviced
Replace o-ring on top of
sampling head.
Replace main seal (o-ring) under
sampling head.
Have unit serviced
Oil leaking from magnet area
Missing o-ring or seal.
Waste collection port blocked.
Waste bottle cap not sealing.
Replace seal and o-ring.
Remove magnet from well and
unblock waste collection port at
bottom.
Tighten waste bottle cap.