Page 13 of 36
6.
Fixer Pump Priming Procedure
Each fixer pump is primed and set to the proper flow rates at the factory before shipment. However, fixer
in the dispensing lines dissipates quickly and will require the operator to re-prime each line before
running Ferrograms on each station when the unit sits for prolonged periods of time.
After the FM-6 is set-up and ready to run, test the fixer pumps by pressing and holding the bottom button
on right and left side of the control panel marked FIXER. If after 60 seconds, the fixer doesn’t flow
follow the procedure below.
1.
Press the DIAGNOSTICS tab which allows the user to turn on and off each of the mechanical
components of the FM-6.
2.
Place a substrate slide onto station one and two.
3.
Select the buttons for Fixer Pulse on station one and station two. This will run the fixer pumps
without holding the FIXER buttons on the RUN FM tab.
4.
Allow fixer to be drawn from the fixer bottle into the pumps and into each line. This may take a
few minutes if both fixer lines are dry.
5.
If no fixer is dispensed, open the fixer compartment door located on the right side of the machine
and remove the fixer bottle.
6.
Using the fixer priming bulb, draw some fixer into the bulb from the fixer bottle.
7.
Gently pull the fixer line out of the fixer compartment.
8.
Insert one of the small tubes from the fixer line into the end of the priming bulb.
9.
Elevate the priming bulb and gently squeeze fixer into the line until the fluid comes out one of the
lines located at the dispensing arm above the magnet assembly.
10.
Turn off the fixer pump in correlation with the fixer line that was primed by pressing either
“Fixer 1 off” or “Fixer 2 off” on the DIAGNOSTICS tab screen.
11.
Repeat the same procedure to prime the second line.
12.
Once primed, insert the fixer line into the fixer bottle and place the bottle into the compartment.
13.
Once fixer starts to flow from the dispensing arms onto the slides turn the waste pump on by
pressing the Waste ON button in the DIAGNOSTICS tab.