Page 18 of 36
8.
Run FM Procedures
8.1 – OIL SAMPLES FOR TESTING
Test samples should be obtained from the equipment based on the predictive maintenance program.
Specific collection techniques and suggested frequencies are suggested below.
•
Always use approved glass or plastic sample bottles.
Never use metal containers.
•
Take samples from the machine while it is operating or no longer than 15 minutes after shut
down. Try to take each successive sample while the machine is operating under the same load.
•
Take samples from the same point on the machine every time.
•
Always take samples upstream from a filter.
•
If using a sample valve, make sure stagnant oil is purged from the assembly prior to filling
sample bottle.
•
Heat sample to 150ºF (65ºC) prior to testing.
•
Special techniques are required to sample pipes and oil tanks. Please refer to the Trico Wear
Atlas for additional information.
8.2 – INITIATING THE FM-6
To initiate the FM-6, follow these basic procedures:
1.
Press the on/off switch on the rear of the FM-6 unit to the on position. The sides of the switch
will indicate red, when the FM-6 is on. When the FM-6 is on Windows 7 will go through its start-
up operation and then the FM program will automatically boot. The RUN FM tab of the FM-6
screen will be shown.
2.
Verify the FM-6 is level prior to running samples. Make adjustments where appropriate using the
leveling feet in front of the unit. Lower the support feet located under the waste and fixer bays for
additional support.
3.
Place a full fixer bottle into the fixer bay located on the right side of the instrument by pressing
the door inward and allowing the door to open. Pull the door away from the magnet closure
assembly. When placing fixer bottle into position, gently put the Viton tubing through the neck of
the bottle and seat bottle into the cradle.
NOTE:
Be sure when positioning the fixer bottle that the label is facing outwards to avoid false
low level readings indicated by the sensor behind the bottle.
4.
Open the waste bay on the left side of the FM-6 and pull the suction cap assembly out.
5.
Thread the waste bottle onto the cap and gently push the assembly into the waste bay.