Incoming Confidence Test (cont.)
8. Continue to move the negatively charged object towards the ion collecting plate
until the value indicated on the PLATE VOLTAGE display passes through the
programmed decay stop voltage.
Note:
The decay stop voltage is factory programmed for 100 V. (See ‘Programming
the STOP voltage’ on page III-5 for information on changing the decay stop
voltage.)
At the programmed stop voltage the decay timer will stop and the decay time will be
indicated on the DECAY TIME display.
This completes the incoming confidence test. Turn OFF the 156A.