Section V Maintenance
Safety Precautions for the Model 156A
Please observe the following safety precautions when checking connections or
providing maintenance procedures on the Model 156A:
1.
Refer all maintenance procedures to qualified personnel.
2.
Prior to performing maintenance procedures, turn off the 156A and disconnect
the battery charger/eliminator from its power source. Failure to observe this
precaution could result in an electrical shock.
3.
Allow a cool-down period to reduce the danger of burns from heated parts such
as transistors and heat sinks.
4.
This instrument incorporates a rechargeable 7.2 V battery pack which is not user
serviceable. Do not substitute a non-rechargeable battery because risk of fire or
explosion could result. Please contact TREK, Inc. or an authorized service
organization for battery replacement information.
Please go to our web site: www.trekinc.com for a complete list of our sales and
service representatives and distributors located in the United States and throughout
the world.
Battery Pack Maintenance
The Model 156A is equipped with a rechargeable battery pack. Periodic cycling
(complete charge and discharge) of the battery pack is required in order to maximize
the life and capacity of the battery pack. Trek, Inc. recommends cycling the battery
pack at least once a month. Cycling of the battery pack can be accomplished by
following the procedure below.
1) Plug the battery eliminator/charger into the Model 156A.
2) Charge the battery pack until the “Full Charge” indicator on the Model 156A is
illuminated.
3) Remove the charger from the Model 156A.
4) Turn the power to the Model 156A on.
V - 1