Section IV Model 156A Specifications
DANGER: This instrument is not rated for use in an explosive environment.
DO NOT use it an explosive environment or an explosion may result.
PERFORMANCE
Monitored Voltage Range
0 to ±1100 V DC or peak AC.
Small Signal Bandwidth (-3 dB)
DC to 1 kHz (measured at 20 V p-p).
Large Signal Bandwidth
DC to 10 Hz (measured at
2000 V p-p).
Zero Stability
(referred to plate voltage)
Drift with Time
(no incident ion flow)
Less than 6 V/minute.
Drift with Temperature
Less than 10 mV/°C, noncumulative.
Decay Mode Thresholds
Start Voltage
Programmable from 1 to
±
1000 V in
1 V increments.
Start Accuracy
Within ±1 V of programmed start
voltage.