Specifications (cont.)
FEATURES (cont.)
Plate Voltage Digital Panel Meter
3 ½ digit red LED display.
Range
0 to ±1100 V.
Resolution
1 volt.
Accuracy
Better than 0.1% of full scale, ±1 count.
Decay Time Digital Panel Meter
4-digit red LED display.
Range
0 to 9999 seconds.
Resolution
0.1 seconds, from 0 to 999.9 seconds;
1 second, from 1000 to 9999 seconds.
(The display will indicate “- - - -” when
the decay time exceeds 9999 seconds.)
Ion Collector Plate
Standard Option
Meets ESD Association Standard
ANSI/ESD-STM3.1 requirements for:
Dimensions
15.3 cm x 15.3 cm square, (6
²
x 6
²
square).
Capacitance
20 pF ±2 pF.
Non-standard Option
Dimensions
Specify when ordering.
Capacitance
Specify when ordering.
Mode Select
A three-position toggle switch that selects
the +Decay, the -Decay, or the Float mode
of operation. This switch is also used in
combination with the Test/Reset Control
switch to program the start and stop
voltages.
IV - 3
Specifications
Содержание 156A
Страница 1: ...1445 JRB OPERATOR S MANUAL MODEL 156A CHARGED PLATE MONITOR 511 1408 400 660 9565 www 19mro com...
Страница 2: ......
Страница 4: ......
Страница 12: ...viii...
Страница 26: ...III 8...
Страница 36: ......
Страница 38: ......
Страница 40: ......
Страница 42: ...511 1408 400 660 9565 www 19mro com...