EMC GUIDANCE

24

 

150 kHz to 80 MHz 

80 MHz to 800 MHz 

800 MHz to 2.5 GHz 

Table 6 Recommended separation distances between portable and mobile 

RF communications equipment and the ME EQUIPMENT or ME SYSTEM –

for ME EQUIPMENT and ME SYSTEMS that are not LIFE-SUPPORTING 

Recommended separation distances 

between  portable and mobile RF communications equipment and the device.

The device is intended for use in an electromagnetic environment in which 

radiated RF disturbances are controlled. The customer or the user of the device

can help prevent electromagnetic interference by maintaining a minimum distance 

between portable and mobile RF communications equipment (transmittters) and the 

device as recommended below, according to the maximum output power of the 

communications equipment.

Rated maximum output

power of transmitter

(W) 

Separation distance according to frequency of transmitter (m)

For transmitters rated at a maximum output power not listed above, the recommended 

separation distance 

d

 in metres (m) can be estimated using the equation applicable to 

the frequency of the transmitter, where 

P

 is the maximum output power rating of the 

transmitter in watts (W) according to the transmitter manufacturer.

NOTE 1    At 80MHz and 800MHz, the separation distance for the higher frequency 

                 range applies.

NOTE 2    These guidelines may not apply in all situations. Electromagnetic propagation

                 is affected by absorption and reflection from structures, objects and people.

0.01

0.1

1

10

100

0.12

0.12
0.38
1.2
3.8
12

0.23
0.73
2.3
7.3
23

=

d

P

=

d

1,2

2,3

P

=

d

0.38
1.2
3.8
12

P

1,2

Содержание TMB-1018-A

Страница 1: ...r selecting TRANSTEK Blood Pressure Monitor TMB 1018 A Please keep this manual well in order to reference in future Arm Type Version 1 0 EC REP Guangdong Transtek Medical Electronics Co Ltd Zone A No...

Страница 2: ...ON FOR USER 13 Tips for measurement Maintenances ABOUT BLOOD PRESSURE 15 What are systolic pressure and diastolic pressure What is the standard blood pressure classification Why does my blood pressure...

Страница 3: ...deflate Should the cuff not deflate when pressures exceeds 40 kPa 300 mmHg detach the cuff from the arm and press the START STOP button to stop inflation The equipment is not AP APG equipment and not...

Страница 4: ...ure 1kPa 7 5mmHg Low battery Batteries are low and need to be replaced Shocking reminding Shocking will result in inaccurate Average The average of blood pressure Recalling The records will be showed...

Страница 5: ...ery cover Install the batteries by matching the correct polarity as shown Replace the cover 1 Battery powered mode 6VDC 4 AAA batteries 2 AC adaptor powered mode 6V 1A Please only use the recommended...

Страница 6: ...he cuff should be snug but not too tight You should be able to insert one finger between the cuff and your arm 3 Sit comfortably with your tested arm resting on a flat surface 2 3cm 4 Repeat steps 2 a...

Страница 7: ...st CAUTION The order of the record date time will display alternately 1 When the monitor is off press the START STOP to turn on the monitor and it will finish the whole measurement Adjust the zero LCD...

Страница 8: ...3 If you don t want to delete the records press START STOP to escape If you did not get the correct measurement you can delete all results by following steps below Delete the Records 4 If there is no...

Страница 9: ...classification The blood pressure classification published by World Health Organization WHO and International Society of Hypertension ISH in 1999 is as follows Irregular Heartbeat Detector CAUTION Th...

Страница 10: ...fer to the warranty for contact information and return instructions E21 shows The treatment of the measurement failed Relax for a moment and then measure again AC adaptor is inserted incorrectly Inser...

Страница 11: ...00 240V 50 60Hz 400mA Output 6V 1A Type UE08WCP 060100SPA Conforms to UL certificate Contact Information For more information about our products please visit www transtek cn you can get customer servi...

Страница 12: ...ablishments and those directly connected to the public low voltage power supply network that supplies buildings used for domestic purposes Risk management Labeling User manual General Requirements for...

Страница 13: ...G Guidance and manufacturer s declaration electromagnetic immunity IMMUNITY test Not applicable 3 V m Compliance level IEC 60601 TEST LEVEL Conducted RF IEC 61000 4 6 Radiated RF IEC 61000 4 3 3 V m 8...

Страница 14: ...the device as recommended below according to the maximum output power of the communications equipment Rated maximum output power of transmitter W Separation distance according to frequency of transmit...

Отзывы: