7.17 Lateral Cassette Holder
Chapter 7: Accessories
Page 7 - 32
TranStar® PC-100/PC-200/PC-300/PC-350/PC-400/PC-450/
PC-500/PC-550/TR-500/SG-500/OB-450/OB-550/EL-550
Series Stretcher Service Manual (man115re)
7.17 Lateral Cassette Holder
The TranStar® Trauma Stretcher Lateral Cassette Holder will rest on the
longitudinal upper frame tube on the side of the stretcher. It allows for side
view x-ray shots of the patient while on the stretcher. It will support a cassette
with a maximum weight of 12 lb (5 kg).
Installation
1. Put the siderails down and in the stored position.
NOTE:
The siderail is not usable when the lateral cassette holder is in position.
2. Grasp the handle (E), and put the lateral cassette holder (B) in its proper
position on the longitudinal upper frame tube (A) (see figure 7-16 on
page 7-32).
Figure 7-16. Lateral Cassette Holder
3. Raise the upper portion of the lateral cassette holder (D) enough to place
the cassette in. Place the cassette on the lateral cassette holder, and adjust
the height if needed using the adjustment knob (C). If necessary, use the
handle (E) to change the location of the lateral cassette holder.
m115b130