Transcell Technology TI-1520 Скачать руководство пользователя страница 8

 

 

CODE/NAME 

DESCRIPTION 

SELECTION LIST 

F5 
Motion Band 

Sets the level at which motion is detected. If motion is not detected, 
the  scale  can  process  a  Print  or  Zero  command.  Maximum  value 
varies depending on local regulations. Expressed as scale divisions 
per  second  (d/s).  Pressing  the  ZERO  key  to  scroll  down  one  level 
begins the sequence 

Key-in  
0.0d/s – 32.0d/s 
 
1.0d/s (NA/€)
 

F6 
Digital Filter 

Averages  weight  readings  to  produce  higher  stability.  Choose  the 
speed that works best for your application.  

   “FAST”  = Fast            “nnEd” = Medium             “SLo” = Slow  

FAST 
nnEd (NA/€) 
SLo 

F7 
Overload Limit 

Selects the desired formula which determines the point at which the 
indicator  shows  overload.  All  selections  are  based  on  the  primary 
unit selected in F8. 

"FS" = Full scale capacity. 

FS 
FS + 2% (NA) 
FS + 5% 
FS + 1d 
FS + 9d (€) 

F8 
Calib. Unit 

Selects the primary base unit to be used in the calibration process. 
Also the default unit for normal operation. 
"1" = primary unit is lb.                       "2" = primary unit is in kg. 

1 (NA) 
2 (€) 

F9 
Display 
Divisions 

Determines  the  desired  weight  increments.  Value  should  be 
consistent with legal requirements.  

1 (NA) 

5 (€) 

F10 
Decimal Pt. 

Determines location of the decimal point.  

0 (NA)      0.0 (€) 
0.00        0.000 
0.0000         00 

F14 
Units 
Conversion 

Allows  the  lb/kg  key  to  be  disabled  so  that  an  operator  cannot 
accidentally press the key and change the displayed units. 

"0" = Disable the Units key              "1" = Enable the Units key 

0 (€) 
1 (NA) 

F16 
Zero 
Calibration 

Places indicator into the zero calibration routine. Scrolling down with 
the ZERO key one level begins the procedure. 

Press ZERO key 
to begin sequence 

F17 
Span 
Calibratio

Places  indicator  into  the  span  calibration  routine.  Scrolling  down 
with the ZERO key one level begins the procedure. 

Press ZERO key 
to begin sequence 

F18 
View 
Calibration 

Actuates the function that allows you to view both the zero and span 
calibration value. The values displayed in this function are valid only 
after  Calibration  (F16  &  F17)  has  been  successfully  completed. 
Scrolling  down  with  the  ZERO  key  one  level  begins  the  procedure. 
Multi-point cal 

Press ZERO key 
to begin sequence 

F19 
Key-in Zero 

Allows you to key-in known zero calibration value in case of memory 
loss in the field. Scrolling down with the ZERO key one level begins 
the procedure. 

Press ZERO key 
to begin sequence 

F20 
Key-in Span 

Allows  you  to  key-in  a  known  span  calibration  value  in  case  of 
memory  loss  in  the  field.  Scrolling  down  with  the  ZERO  key  one 
level begins the procedure. 

Press ZERO key 
to begin sequence 

 

Содержание TI-1520

Страница 1: ...MODEL TI 1520 Digital Weight Indicator User Manual Revision 1 8 August 3 2012...

Страница 2: ...EW CALIBRATION VALUES F18 12 KEY IN ZERO CALIBRATION VALUE F19 13 KEY IN SPAN CALIBRATION VALUE F20 13 OPERATION 14 DISPLAY 14 KEYBOARD 15 FUNCTION KEYS 15 GENERAL SCALE OPERATION 15 WEIGHING AN ITEM...

Страница 3: ...idential installation This equipment generates uses and can radiate radio frequency energy and if not installed and used in accordance with the instructions may cause harmful interference to radio com...

Страница 4: ...ssible In extreme cases it may be necessary to install surge suppressors line conditioners or even UPS Uninterruptible Power Supplies systems not included Environment Avoid installing the indicator in...

Страница 5: ...o the green load cell input terminal The TI 1520 indicator contains a connection terminal on the main board for connection to the load cell cable Connect your shielded load cell cable not included to...

Страница 6: ...The Setup and User menus consist of several menu selections each with its own sub menu of selections or programming procedures To configure the indicator you must first enter the appropriate menu mode...

Страница 7: ...S This section provides more detailed descriptions of the selections found in the Setup Menu Chart Factory set defaults are shown in bold NA for North America and for Europe NOTE Some selections are s...

Страница 8: ...al point 0 NA 0 0 0 00 0 000 0 0000 00 F14 Units Conversion Allows the lb kg key to be disabled so that an operator cannot accidentally press the key and change the displayed units 0 Disable the Units...

Страница 9: ...0 0 NA 10 F32 Center of Zero Band Selects the range around gross zero within which the scale will display the Center of Zero annunciator Selections are in display divisions d 0 25d US 0 5d F34 Auto Pr...

Страница 10: ...Loads in a pre defined output string print format for the serial port Refer to Appendix B for details 0 Text Print Ticket 1 String Format 1 Condec Demand 2 String Format 2 Condec Continuous 3 Text Pri...

Страница 11: ...alue Increase the flashing digit by pressing the UNITS key Decrease the flashing digit by pressing the ZERO key Pressing the PRINT key or the TARE key will change the position of the flashing digit 4...

Страница 12: ...3 Use the four directional keys to adjust the displayed value to the actual date value Increase the flashing digit by pressing the UNITS key Decrease the flashing digit by pressing the ZERO key Press...

Страница 13: ...troubleshoot setup problems 2 After making sure that there are no test weights on the platform press the ZERO key again to zero out the displayed value 3 Press the NET GROSS key to save the zero point...

Страница 14: ...causes include incorrect load cell wiring a mechanical obstruction or a faulty load cell VIEW CALIBRATION VALUES F18 Note The values displayed in this procedure are valid only after a successful cali...

Страница 15: ...id span calibration values obtained from a successful F17 calibration procedure must be used 1 While in the Setup mode scroll to F 20 and then scroll down once using the ZERO key The indicator will pr...

Страница 16: ...ever the displayed weight is within a pre programmed band from true zero NET Indicates that the indicator is displaying net weight GROSS Indicates that the indicator is displaying gross weight TARE In...

Страница 17: ...scale is not in overload see Appendix D for error codes Print This key is used to send weight information out to the serial port provided the following conditions are met 1 The scale is not in motion...

Страница 18: ...u continue to push the UNITS key the indicator will resort back to weighing mode and you must start again from Step 2 3 Place the sample items on the platform all at once and allow the weight indicati...

Страница 19: ...20 continuously and at the same transmission baud rate configured in A1 LEGAL FOR TRADE SEALING Indicators can be sealed for commercial Legal for Trade applications as follows 1 On the back of the ind...

Страница 20: ...56 x 8 external to C Digital Filtering Software selectable SERIAL COMMUNICATIONS Serial Port Full Duplex selectable Baud rate 8 data bits no parity 1 stop bit 7 data bits odd parity 1 stop bit 7 data...

Страница 21: ...tor to zero the scale The indicator will not respond if the scale is in motion positive overload or negative overload The indicator will also not respond if it is in net mode or outside the zero range...

Страница 22: ...ht TWT 131 83 Print the data on screen CWT 132 85 Units lb kg PCS UN 133 86 Date DAT 134 87 Time TIM 135 88 O Over Under M Motion STAT 136 89 Gross Net flag G N GN 137 8A MP20 printer to print header...

Страница 23: ...02 2E 31 53 S 03 4E N 32 80 GWT 04 4F O 33 85 UN 05 2E 34 0D CR 06 8C PRID 35 0A LF 07 0D CR 36 54 T 08 0A LF 37 41 A 09 44 D 38 52 R 10 41 A 39 45 E 11 54 T 40 20 12 45 E 41 82 TWT 13 8E PRSPS 42 85...

Страница 24: ...larity SP Positive Negative Weight Data Units LB pound KG kilogram Gross Net GR Gross NT Net Start Transmission Carriage Return Line Feed SP SP Space Space pc pieces cu cusotm unit pcs pieces Address...

Страница 25: ...F Polarity SP Positive Negative Weight Data Units L pound K kilogram Gross Net G Gross N Net Start Transmission Carriage Return Line Feed Status SP Valid M Motion O Over under range P pieces C custom...

Страница 26: ...rrow key PRINT to increase the address by one Use the left arrow key TARE to decrease the address by one 3 After selecting the address you want to change press the NET GROSS key The programming value...

Страница 27: ...Keyed in weight value is less than 1 of full scale capacity Use a larger test weight or check keyed in value Err 2 Span Calibration Mode F17 There is not enough load cell signal to produce the intern...

Отзывы: