Percentage Operation

 

 

The EL 3000 Digital Scale contains a percentage operation which is useful for such applications as 
filling, check weighing and moisture determination. 

The total weight of the item being sampled determines the resolution of the percentage readout. For 
example, if the sample weight is greater than or equal to 10,000 external scale divisions, the scale will 
discern 0.01% increments. See table below more information. 

To use the percentage operation: 

1. Toggle 

the 

MODE 

key until the scale is in “%” mode. If there is no previous sample entered, the 

scale will respond with the following display indication. If the following message does not appear, 
press the 

SAMPLE

 key once. 

This display indication is prompting you to 
place 100% of the sample on the platform all 
at once. 

NOTE

: If you entered this screen by 

accident, you can press the 

MODE

 key to 

escape. 

G

oz

T

N

lb

pcs

%

OK

HI

LO

O

E

kg

 

2. Press 

the 

ZERO

 key to zero the scale. 

3.  Place the sample on the platform all at once. 

When the "Stable" flag becomes lit, the 
display should look like this: 

G

oz

T

N

pcs

%

OK

HI

LO

O

E

lb
kg

 

4. Press 

the 

SAMPLE

 key. If the sample weight is over the required sample weight, the scale will now 

show the sample as being 100%. If not, a “Lo” message will flash briefly, prompting you for a 
heavier sample. See table below for more information. 

Percentage Operation Sample Weight Guide: 

 

The minimum sample weight is 100 scale graduations. This will allow the scale to discern 1 part in 100. 
To get a higher resolution, use 1,000 or even 10,000 times the scale’s graduations as your sample 
weight. Below are three examples to help determine the best sample weight for your application. 

Capacity / Graduation 

Minimum Sample Weight 

(100 Graduations) 

Best Sample Weight  

(10,000 Graduations) 

30 lbs x 0.002-lbs 

0.200-lbs 

20-lbs 

1200 g x 0.1 g 

10 g 

1000 g 

6 lbs x 0.0005-lbs 

0.05-lbs 

5-lbs 

 

 

 

EL 3000 Operating Manual

   

Содержание EL 3000

Страница 1: ...1 April 16 2002 1995 2002 Transcell Technology Inc Contents subject to change without notice Transcell Technology Inc 975 Deerfield Parkway Buffalo Grove IL 60089 Tel 847 419 9180 Fax 847 419 1515 E m...

Страница 2: ...warranty Table of Contents Introduction 1 Getting Started 1 Basics of Operation 2 Percentage Operation 3 Piece Counting Operation 4 Accumulator Function Operation 5 Disabling or Enabling Units 6 User...

Страница 3: ...r unpacking the scale a small amount of assembly is required Locate the sub platform four 4 Hex screws and enclosed Allen wrench Place sub platform onto scale s load point with all rubber parts facing...

Страница 4: ...ntage modes Low Battery Indication INDICATION MEANING Weight Indicates the weight currently on the platform When the full scale capacity has been exceeded the scale shows Units Indicates the unit of t...

Страница 5: ...the MODE key to escape G oz T N lb pcs OK HI LO O E kg 2 Press the ZERO key to zero the scale 3 Place the sample on the platform all at once When the Stable flag becomes lit the display should look li...

Страница 6: ...ication If the following message does not appear press the SAMPLE key once This display indication is prompting you to place 10 identical samples on the platform all at once NOTE If you entered this s...

Страница 7: ...MODE key to toggle to the desired unit you wish to count 2 If the MEM indication is lit press and hold the ZERO CLEAR key to clear the accumulator memory The scale briefly displays CL to indicate tha...

Страница 8: ...tch on the bottom of the scale near the back 2 While holding down the MODE key turn the scale back on Continue holding down the MODE key until the scale shows the LB unit with either the ON message or...

Страница 9: ...ave or SET the selections Complete directions start below To place the unit in USER MENU mode Turn the scale off While holding down the MR key turn the scale back on When the display shows A1 the unit...

Страница 10: ...utes Scale must be idle during this period OFF Indicates this feature is disabled OFF 1 3 5 10 15 20 30 NAME CODE DESCRIPTION CODE VALUE Notes on A5 Auto Power Off Period 1 Should be set to OFF if the...

Страница 11: ...cline in the usage period This is due to the bat tery s reluctance to accept a charge Therefore always charge the scale for at least 10 hours before using 3 The charger may be left connected to the sc...

Страница 12: ...ghtmost digit should be flashing Place the test weight on the platform 4 Use the four directional keys shown above to adjust the displayed value to the actual test weight value Increase the flashing d...

Страница 13: ...erface to printers the Demand mode allows control from a host device usually a PC and can be activated by pressing the MR key on the front panel Full Duplex Continuous Mode XXXX XXX UNITS S1 ETX Polar...

Страница 14: ...ogy Demand Mode Valid Weight Reading 1 Recognized host command s P for weight Data Z to zero the scale C to change displayed units S1 ETX End Transmission Status Character 1 See Continuous Mode Questi...

Страница 15: ...cid battery Charger AC Adapter 15 VDC 800mA Charge Period 8 10 hours WEIGHT Net Weight 8 8 lbs 4 kg Shipping Weight 14 2 lb 6 4 kg Battery Option Add 1 2 lb 0 5 kg SERIAL PORT Full Duplex RS 232 forma...

Страница 16: ...affect stability Scale should be located away from air conditioners fans fume hoods etc A breeze shield may be required D Extreme changes in room temperature can affect the internal calibration of th...

Страница 17: ...exists and was not caused by accident misuse neglect alteration improper installation improper or un authorized repair or improper testing and 4 Such goods have not been modified altered or changed b...

Отзывы: