In Case of Trouble... 

 

After reading your operating manual thoroughly, you should have no trouble using your 
scale. However, if you are still experiencing difficulty, please review the following points 
before calling or returning the scale. 
 

PROBLEM

#1.  Unstable Reading
#2.  Does Not  Weigh
#3.  No Display

X

X

X

X

X

X

X

X

X

X
X

X

 

 

PROBLEM: 

1.  When sample is placed on pan, display 

reading will not stabilize, i.e. jumps one 
or more divisions. 

2.  Display will not change after sample is 

placed on platform or always shows 
overweight condition. 

3.  No readout appears in the display win-

dow. 

 

THINGS TO CHECK: 

A. Proper warm-up time is at least 30 

minutes. It is recommended that the unit 
be left on continuously. 

B.  Large machinery may cause vibration in 

the work area. The scale should be 
placed on a stable surface. 

C. Excessive air movement can affect 

stability. Scale should be located away 
from air conditioners, fans, fume hoods, 
etc. A breeze shield may be required. 

D.  Extreme changes in room temperature 

can affect the internal calibration of the 
unit. 

E. The sub-platform is held in place by four 

screws which may vibrate loose. Re-
move the platform to check. 

IMPOR-

TANT: UNPLUG THE AC ADAPTER 
FROM THE AC OUTLET PRIOR TO 
REMOVING THE PLATFORM. Do not 
over tighten screws.

 

F.  Do not place the scale near magnetic 

materials or instruments which incorpo-
rate magnets in their design. 

G. Replace AC Adapter. 

IMPORTANT: 

Replace only with same voltage, 
current, and polarity of originally 
supplied adapter.

 

H.  The line voltage to the scale should be 

reasonably constant and free from fluc-
tuations. It is advisable not to share an 
outlet with fluorescent lights or other 
products which draw voltage in an in-
consistent manner. 

I.  Calibration memory has been lost. Re-

calibrate scale. 

J.  A/D converter is not stable or load cell 

has failed. Return scale for repair. 

14

   

Revision 2.95 

Содержание EL 3000

Страница 1: ...1 April 16 2002 1995 2002 Transcell Technology Inc Contents subject to change without notice Transcell Technology Inc 975 Deerfield Parkway Buffalo Grove IL 60089 Tel 847 419 9180 Fax 847 419 1515 E m...

Страница 2: ...warranty Table of Contents Introduction 1 Getting Started 1 Basics of Operation 2 Percentage Operation 3 Piece Counting Operation 4 Accumulator Function Operation 5 Disabling or Enabling Units 6 User...

Страница 3: ...r unpacking the scale a small amount of assembly is required Locate the sub platform four 4 Hex screws and enclosed Allen wrench Place sub platform onto scale s load point with all rubber parts facing...

Страница 4: ...ntage modes Low Battery Indication INDICATION MEANING Weight Indicates the weight currently on the platform When the full scale capacity has been exceeded the scale shows Units Indicates the unit of t...

Страница 5: ...the MODE key to escape G oz T N lb pcs OK HI LO O E kg 2 Press the ZERO key to zero the scale 3 Place the sample on the platform all at once When the Stable flag becomes lit the display should look li...

Страница 6: ...ication If the following message does not appear press the SAMPLE key once This display indication is prompting you to place 10 identical samples on the platform all at once NOTE If you entered this s...

Страница 7: ...MODE key to toggle to the desired unit you wish to count 2 If the MEM indication is lit press and hold the ZERO CLEAR key to clear the accumulator memory The scale briefly displays CL to indicate tha...

Страница 8: ...tch on the bottom of the scale near the back 2 While holding down the MODE key turn the scale back on Continue holding down the MODE key until the scale shows the LB unit with either the ON message or...

Страница 9: ...ave or SET the selections Complete directions start below To place the unit in USER MENU mode Turn the scale off While holding down the MR key turn the scale back on When the display shows A1 the unit...

Страница 10: ...utes Scale must be idle during this period OFF Indicates this feature is disabled OFF 1 3 5 10 15 20 30 NAME CODE DESCRIPTION CODE VALUE Notes on A5 Auto Power Off Period 1 Should be set to OFF if the...

Страница 11: ...cline in the usage period This is due to the bat tery s reluctance to accept a charge Therefore always charge the scale for at least 10 hours before using 3 The charger may be left connected to the sc...

Страница 12: ...ghtmost digit should be flashing Place the test weight on the platform 4 Use the four directional keys shown above to adjust the displayed value to the actual test weight value Increase the flashing d...

Страница 13: ...erface to printers the Demand mode allows control from a host device usually a PC and can be activated by pressing the MR key on the front panel Full Duplex Continuous Mode XXXX XXX UNITS S1 ETX Polar...

Страница 14: ...ogy Demand Mode Valid Weight Reading 1 Recognized host command s P for weight Data Z to zero the scale C to change displayed units S1 ETX End Transmission Status Character 1 See Continuous Mode Questi...

Страница 15: ...cid battery Charger AC Adapter 15 VDC 800mA Charge Period 8 10 hours WEIGHT Net Weight 8 8 lbs 4 kg Shipping Weight 14 2 lb 6 4 kg Battery Option Add 1 2 lb 0 5 kg SERIAL PORT Full Duplex RS 232 forma...

Страница 16: ...affect stability Scale should be located away from air conditioners fans fume hoods etc A breeze shield may be required D Extreme changes in room temperature can affect the internal calibration of th...

Страница 17: ...exists and was not caused by accident misuse neglect alteration improper installation improper or un authorized repair or improper testing and 4 Such goods have not been modified altered or changed b...

Отзывы: