Optional Battery Operation

 

 

Here are a few notes for those users who ordered the Battery Operation option for the EL-3000: 

 

Before using the scale for the first time, please charge the scale overnight. 

 

The scale’s battery should operate for about 10 hours if left on continuously. Therefore, greater 
usage times can be achieved by selecting an appropriate Auto Power Off Period under 

A5

 of the 

User Menu. 

 

The scale can be charged while ON or OFF and can be operated while charging. 

To charge the internal battery: 

1.  Connect the charger (AC Adapter 15 VDC, 800mA) to the scale , then plug the charger into an AC 

outlet. 

2.  After the charging period expires, unplug the charger from the AC outlet, then from the scale. The 

scale is now ready for use under its own battery power. If the scale is not allowed for charge for the 
full recommended time period, you may notice a decline in the usage period. This is due to the bat-
tery’s reluctance to accept a charge. 

Therefore, always charge the scale for at least 10 hours 

before using.

 

3.  The charger may be left connected to the scale indefinitely without damage to the internal battery. 

When to charge the internal battery: 

1.  When the scale needs to be charged, the Low Battery Indicator will appear in the lower right-hand 

corner of the display. The scale may be used for an additional 15 to 30 minutes without damage to 
the internal battery. 

If the scale is used for more than 30 minutes after a low battery 

indication, permanent damage may occur to the battery and/or the battery may not accept a 
charge. 

 

2.  With the sealed lead-acid type battery, it is OK to charge the battery at any time to keep the battery 

“fresh”. You do not necessarily have to wait for the Low Battery Indication. 

NOTE:

 When a low battery indication occurs, the scale automatically shuts off after 5 minutes if the 

scale is idle. The period selected for Auto Power Off Period (A5) is not altered. 

 

 

EL 3000 Operating Manual

   

Содержание EL 3000

Страница 1: ...1 April 16 2002 1995 2002 Transcell Technology Inc Contents subject to change without notice Transcell Technology Inc 975 Deerfield Parkway Buffalo Grove IL 60089 Tel 847 419 9180 Fax 847 419 1515 E m...

Страница 2: ...warranty Table of Contents Introduction 1 Getting Started 1 Basics of Operation 2 Percentage Operation 3 Piece Counting Operation 4 Accumulator Function Operation 5 Disabling or Enabling Units 6 User...

Страница 3: ...r unpacking the scale a small amount of assembly is required Locate the sub platform four 4 Hex screws and enclosed Allen wrench Place sub platform onto scale s load point with all rubber parts facing...

Страница 4: ...ntage modes Low Battery Indication INDICATION MEANING Weight Indicates the weight currently on the platform When the full scale capacity has been exceeded the scale shows Units Indicates the unit of t...

Страница 5: ...the MODE key to escape G oz T N lb pcs OK HI LO O E kg 2 Press the ZERO key to zero the scale 3 Place the sample on the platform all at once When the Stable flag becomes lit the display should look li...

Страница 6: ...ication If the following message does not appear press the SAMPLE key once This display indication is prompting you to place 10 identical samples on the platform all at once NOTE If you entered this s...

Страница 7: ...MODE key to toggle to the desired unit you wish to count 2 If the MEM indication is lit press and hold the ZERO CLEAR key to clear the accumulator memory The scale briefly displays CL to indicate tha...

Страница 8: ...tch on the bottom of the scale near the back 2 While holding down the MODE key turn the scale back on Continue holding down the MODE key until the scale shows the LB unit with either the ON message or...

Страница 9: ...ave or SET the selections Complete directions start below To place the unit in USER MENU mode Turn the scale off While holding down the MR key turn the scale back on When the display shows A1 the unit...

Страница 10: ...utes Scale must be idle during this period OFF Indicates this feature is disabled OFF 1 3 5 10 15 20 30 NAME CODE DESCRIPTION CODE VALUE Notes on A5 Auto Power Off Period 1 Should be set to OFF if the...

Страница 11: ...cline in the usage period This is due to the bat tery s reluctance to accept a charge Therefore always charge the scale for at least 10 hours before using 3 The charger may be left connected to the sc...

Страница 12: ...ghtmost digit should be flashing Place the test weight on the platform 4 Use the four directional keys shown above to adjust the displayed value to the actual test weight value Increase the flashing d...

Страница 13: ...erface to printers the Demand mode allows control from a host device usually a PC and can be activated by pressing the MR key on the front panel Full Duplex Continuous Mode XXXX XXX UNITS S1 ETX Polar...

Страница 14: ...ogy Demand Mode Valid Weight Reading 1 Recognized host command s P for weight Data Z to zero the scale C to change displayed units S1 ETX End Transmission Status Character 1 See Continuous Mode Questi...

Страница 15: ...cid battery Charger AC Adapter 15 VDC 800mA Charge Period 8 10 hours WEIGHT Net Weight 8 8 lbs 4 kg Shipping Weight 14 2 lb 6 4 kg Battery Option Add 1 2 lb 0 5 kg SERIAL PORT Full Duplex RS 232 forma...

Страница 16: ...affect stability Scale should be located away from air conditioners fans fume hoods etc A breeze shield may be required D Extreme changes in room temperature can affect the internal calibration of th...

Страница 17: ...exists and was not caused by accident misuse neglect alteration improper installation improper or un authorized repair or improper testing and 4 Such goods have not been modified altered or changed b...

Отзывы: