3.2 Getting started
Instructions for membrane filtration are covered in another section of the kit manual.
The user should familiarise themselves fully with the set-up, control and calibration of the
incubator prior to carrying out tests for the first time.
If the incubator is connected to the docking station turn on using the switch on the rear of the
carry case. If connected directly to the power supply turn on the power at the AC mains outlet.
The incubator screen will turn on.
The incubator may be in one of three operating modes on power up:
1: New run
2: Partial incubation cycle (Idle)
3: Partial incubation cycle (Run)
If the incubator is in mode 1 or 2 the screen shown in Fig 1.0 (see page 8) will be displayed. If the
incubator is in mode 3 the incubator will return to the last know state and auto start the
incubation cycle from where it was before the power was removed. Mode 3 is primarily intended
for unattended power failure conditions.
If Fig 1.0 is displayed and the incubator was in mode 2 then the previous incubation cycle can be
restarted from where it was left by pressing restart. This will be discussed later.
The screen will be displayed for 60 seconds before defaulting to the restart condition.
Pressing the right hand switch will reset the incubator ready to set-up and start a new incubation
cycle and the screen will change to the one shown in Fig 1.1 (see page 8).
The incubator has two independently controlled thermal chambers which can be configured and
calibrated independently.
The left hand switch primarily selects the mode for the left hand chamber and the right hand
switch the mode for the right hand chamber. The centre switch will perform selections for each
side which will be indicated by an arrow pointing left or right < >.
Each screen and options will now be discussed in further detail.
Содержание AquaSafe WSL50
Страница 1: ...AquaSafe WSL50 Water Safety Laboratory Instruction Manual...
Страница 4: ...Section 2 Kit Contents 1 2 3 4 5 5 Under Accessories Box 6 7 8 a b c d e f g h i j k k l b b a b...
Страница 7: ...Section 3 AquaSafe Micro biological Incubator Instruction Manual...
Страница 28: ...Section 4 AquaSafe Membrane Filtration Instruction Manual Contents...
Страница 44: ...Section 5 AquaSafe Turbidity Tube Instruction Manual...
Страница 47: ...Section 6 AquaSafe Block Tester Instruction Manual for WSL Range...